Operating instructions

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 43
Servicing Instructions 227 436 37 Rev. B V1.3
Printer" wizard. Be sure to select a local LPT port. When you
see the list of manufacturers and models, click the "Have Disk"
button. The necessary files are in the root directory of the
recorder software distribution diskette. (You will also be
prompted for the original Windows NT 4.0 distribution CD.)
Select the "AR200P 8 inch Recorder," and follow the remaining
prompts. Choose not to print a test page.
When the printer driver installation is complete, "GSI AR200P
8 inch Recorder" should appear on the list of printers. Right
click on it and select "Document Defaults" from the pop-up menu.
Click on the "Advanced" tab of the property sheet. Under
"Document Options," click on "Metafile Spooling" and turn it off.
Turn "Print Text as Graphics" on, and verify that "Print Quality"
is set to normal.
Under "Printer", "Properties", "Scheduling" click on:
"Print directly to Printer".
This chapter is only relevant, if you do not have installed CardioSoft.
CardioSoft uses the Default Port LPT1.
The next step in the AR200P installation procedure is to copy the
AR200.INI file from the root directory of the distribution
diskette to the computer's "Windows" directory (often C:\WINNT).
If you have elected to use LPT2 or LPT3 instead of the default
setting of LPT1, use an ascii text editor on the copy of the
AR200.INI file in the "Windows" directory to change the "port=1"
line to "port=2" or "port=3" as appropriate.
Next, create an AR200P directory on your system. Copy all of the
files in the AR200P directory on the distribution diskette to the
new directory.
Finally, run the ARDEMO program to verify that the recorder and
software are operating correctly, and to determine the version
number of the firmware resident in the AR200P flash eprom memory.
To see the version number, click the "Help" menu choice on
ARDEMO's menu bar, and then click "About." The item of interest
is the firmware "BIC" version number.
If the "BIC" firmware version number is less than 1.35, it must
be updated by running the ARUPDATE utility program supplied on
the distribution diskette. ARUPDATE WILL NOT RUN UNDER WINDOWS
NT. YOU MUST USE WINDOWS 3.1 OR MSDOS. See the ARUPDATE.TXT
file for instructions.