Operating instructions
Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 31
Servicing Instructions 227 436 37 Rev. B V1.3
// Some printer drivers have problems managing the printout area, so that the last line //
may be missing, for example. Where printing should start can be set as follows:
PRI_FrameLeft=? // left-hand margin in 1/10 mm
PRI_FrameTop=? // top margin in 1/10 mm
5.2.7.2 Settings for Russian under Windows 98
GRA_ScreenFont2=”Arial Cyr”
GRA_ScreenFont3=”Courier New Cyr”
GRA_PrinterFont1=”Times New Roman Cyr”
GRA_PrinterFont3=”Courier New Cyr”
GRA_PrinterFont5=”Times New Roman Cyr”
5.2.8 Entries in AR200.INI
The DLL for the General Scanning electrothermal printer AR200 accesses this
initialization file. During installation it is copied into the Windows directory.
StartWithTopOfForm=? // Range of values: 0..1 Default: 0
// Meaning: At the beginning of an examination
the paper is fed the beginning of the
page
StopWithFormFeed=? // Range of values: 0..1 Default: 0
// Meaning: At the end of an examination the
paper is fed approx. 3 cm above the
beginning of the next page
The setting “StartWithTopOfForm=1"
must be selected so that the next
examination recording can be made
starting at the beginning of a page.
StopWithTopOfForm=? // Range of values: 0..1 Default: 0
// Meaning: At the end of an examination the
paper is fed to the beginning of the next
page
EcgSize=? // Range of values: 0..4 Default: 3
// Meaning: Thickness of trace printed out during
a recording
1 : very thin
2 : thin
3 : normal
4 : thick
// only takes effect while recording