Operating instructions

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 30
Servicing Instructions 227 436 37 Rev. B V1.3
GRA_CharSet=? // Range of values: 0..4 Default: 0
// Meaning: Determines the character set for the
CardioSoft fonts.
0: ANSI_CHARSET
1: DEFAULT_CHARSET
2 :SYMBOL_CHARSET
3: SHIFTJIS_CHARSET
4: OEM_CHARSET
// Recommendation for Japanese:
GRA_ScreenFont Factor=? // Range of values: 50..500 Default:100
// Meaning: Multiplication factor for screen display
fonts
RES_H1_Format1=V1,V2,V3,V4,V5,V6 //6 leads for 1. H1 format (separated by decimal
point)
RES_H1_Format2=I,II,III,V1,V2,V3 //6 leads for 2. H1 format (separated by decimal
point)
5.2.7.1 Setting 12SL analysis
12SL_Measurement=? // Range of values: 0..1 Default: 0
// Meaning: 0: analysis with “Hellige Heart”
1: analysis with “marquette 12SL”
// additional settings possible for the 12SL analysis
12SL_SCREEN_CRIT=? // Range of values: 0..1 Default: 0
// Meaning: when 1 has been selected the following
12SL findings do not appear on the ECG reports:
ANTERIOR INFARCTION
INFERIOR INFARCTION
NONSPECIFIC ST ABNORMALITY
NONSPECIFIC T-WAVE ABNORMALITY =
suppressed only when the abnormality is
ABNORMAL QRS T ANGLE, CONSIDER T-WAVE
ABNORMALITY
12SL_SUPPR_NORMAL=? // Range of values: 0..1 Default: 0
// Meaning: when 1 has been selected the 12SL
finding NORMAL ECG is prevented from appearing
in the ECG report.
12SL_SUPPR_ABNORMAL=? // Range of values: 0..1 Default: 0
// Meaning: when 1 has been selected the 12SL
finding BORDERLINE FINDING or ABNORMAL
ECG is prevented from appearing in the ECG
report.
12SL_SUPPR_DIAG=? // Range of values: 0..1 Default: 0
// Meaning: when 1 has been selected this
prevents a 12SL diagnosis being printed out.