Operating instructions
Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 29
Servicing Instructions 227 436 37 Rev. B V1.3
10. The stage names are fixed as the following table shows:
Phase Stagenumber Stagename in the shared file (4 characters)
Pre-Test 1 base
Pre-Test 2 warm
... ... warm
Exercise 1 ramp
... ... ramp
Recovery 1 reco
... ... reco
[EGM_RECORD]
SpiroOption=1 //1: initiate output
//0 or not available: no output
SpiroDirectory=C:\SPIROAPP\DATA // default: C:\TEMP
[CARDIO]
CFG_ProdName=GE marquette CardioSys // Poduct name for CardioSys
GE marquette MicroLab // Poduct name for MicroLab
CFG_CorinaTime=5
CFG_Abcux=2977 // 2978 or entry not available: USA version
// 2977: international version
If the file win.ini is inadvertently deleted, only the USA version of CardioSoft can be
started. It is thus advisable to make a backup of the INI files from time to time.
CFG_StressAndRest= // 1: USA version also has resting ECG option
RES_ShowQTDispTable=? // Range of values: 0..1 Default: 0
// Meaning: Output of the QT data for all leads in the
measurement table
GRA_PrinterFontFactor=? // Range of values: 50..500 Default: 100
// Meaning: Font size factor for print fonts
GRA_ScreenFactor=? // Range of values: 50..500 Default: 100
// Meaning: Multiplication factor for screen display
fonts
GRA_GridWeight=? // Range of values: 1..9 Default: 5
// Meaning: Grid points in printouts
PRI_EcgSize=? // Range of values: 0..5 Default: 2
// Meaning: Thickness of the ECG waveform traces
on printouts
// 0 and 1 might cause gaps when using an AR200