Operating instructions
Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 28
Servicing Instructions 227 436 37 Rev. B V1.3
CardioSys updates every 5 seconds the shared file with the following string:
[recoxxx34xMxxxx0x%xxx60xxxx4x-999x-999x-999xx-19xxxx1xx-12x]4CCR
where a space is shown as an 'x'.
Part of string Meaning Notes
[ Beginning
reco Stage 'Recovery' 5, 10
xxx34 Speed * 10 3.4 5, 8
xM Speed designator MPH 1
xxxx0 Grade * 10 0.0 5
x% Grade designator % 2
xxx60 Heart Rate 60 BPM 5
xxxx4 Ectopic Beats per Minute 4 3, 6
x-999 BP Measurement NBR -999 6, 9
x-999 Systolic BP -999 4, 6
x-999 Diastolic BP -999 4, 6
xx-19 ST Level [mm] * 10 -1.9 6
xxxx1 ST Slope [mV/s] * 10 0.1 6
xx-12 ST Integral uV -12 6, 9
x] Ending
4C Checksum 7
CR Carriage return
Notes:
1. Could be 'K' for KPH. Kilometers per hour. If so, SPEED is KPH.
Could be 'P' for KPM. Kilopondmeters. If so, SPEED is KPM from Ergo.
Could be 'W' for WATTS. Watts. From Ergo.
2. Could be 'D' dor DEGREES. If so, GRADE is DEGREES.
3. This is ECTOPIC COUNT for the minute previous.
4. mmHg
5. If unavailable, pass 0.
6. If unavailable, pass -999.
7. Rightmost 2 ASCII characters of checksum. Checksum includes
[, all spaces,] Alpha characters must be Upper Case.
8. If Ergometer Watts or KPM, SPEED value is not *10.
9. Not supported in the current version (value = -999)