Operating instructions
Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 26
Servicing Instructions 227 436 37 Rev. B V1.3
Installation Steps:
1. Install CardioSoft V4.1 on the PC
2. Start the CardioSoft Application and view the System Configuration. In the Tab
Modem select the Baud Rate and the Modem Type (The Port Setting is not used for
the Communication Server). In the Tab MUSE you can configure to send the Rest
ECG's to MUSE (The setting "Start Modem connection before transfer" cannot be
used). In the Tab General please deactivate the "Enable Password Function".
3. For every Modem create a Shortcut to the CardioSoft Application. In the Properties
Dialog of the Shortcut view Tab Shortcut. For Modem 1 at COM1 change the content
of the field "Target" to "...\cardio.exe cartmodem com1" and rename the shortcut to
"CardioSoft Modem 1". Change the other Shortcuts in the same way.
4. Connect all Modems and switch on all Modems. Start the first Shortcut. It should be
displayed a window with the title "COM1: Data received form ECG System/ECG
recorder". If a error message is displayed check the Modem Settings. Press the "End"
Button in the window and close the CardioSoft Application. Test every shortcut in the
same way.
5. Copy all Shortcuts to the Menu Group "Start up" with the Windows NT Explorer.
6. Close all applications and reboot the system.
With every start up to 4 CardioSoft Applications are started and store the received Rest
ECG's. If errors messages are displayed, than check the error like described in
Position 4.
5.2.6 General Entries in *.INI Files, CA_SYS.INI
Procedure Setups will be stored in CARDIO.INI. General settings will be stored in
CA_SYS.INI and could be exported to floppy disc. Special service settings are stored in
WIN.INI (see 5.2.7). When CardioSoft is started the first time and there is no
CA_SYS.INI or no entry in this INI file the default values from the WIN.INI will be used
and copied to the CA_SYS.INI file. After that this values from the CA_SYS.INI file are
used.
5.2.7 Entries in WIN.INI
Entries from the file WIN.INI described here are mainly those which cannot be changed
directly by changing settings in CardioSoft.
Font entries, full screen, HF output ...
[CARDIO]
GRA_ScreenFont1 = “...” // Default =”MS Sans Serif”
GRA_ScreenFont2 = “...” // Default =”Arial”
GRA_ScreenFont3 = “...” // Default =”Courier New”
GRA_ScreenFont4 = “...” // Default =”Small Fonts”
GRA_ScreenFont5 = “...” // Default =”MS Sans Serif”