Operating instructions
Marquette Hellige GmbH CORINA Page 9
Version 2.0 227 437 01 D
2. Introduction
The “Hardware Design Description“ describes the structure of the hardware implemented,
internal interfaces and those for connection of peripherals.
2.1. Repair Procedure
CORINA can be repaired only with the standard repair kit.
A number is stored in the CORINA which allows the accurate assignment of user and soft-
ware options.
Repairs conducted by:
- Availability of AT PCBs CORINA (4 types, 2 with and 2 without the possibility of
plugging in the PCB CORINA TRIGGER)
- Availability of a programming tool, i.e., software in English. The programming tool is a
part of CardioSoft. (...\CARDIO\SERIEN.EXE)
This tool enables the authorized service technician to program the replaced CORINA PCB
with the customer-specific serial number. The AT PCB is programmed with a specific
number. This number only allows the PCB to be modified by overwriting.
Replacement works with the AT PCB only. The AT PCB can be programmed once only.
To come into the programming mode you have at first to click to: FILE --> LOGIN
Password = HELSERV
Then click to: “CORINA --> Set CORINA serial number”
For typing in the serial no, please use only the last 7 digits from the serial no on your name
plate of the CORINA! (With CardioSys or MicroLab please use the serial no. of the
complete system!)
For example: CORINA Serial No: 101 003 789
Type in only: 1 003 789
are not used
No repair kits are necessary for models 5 + 6, they do not contain a serial number. If nec-
essary, as in the case of PCB CORINA TRIGGER, the PCB CORINA can be replaced.