Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 25
Servicing Instructions 227 436 37 Rev. B V1.3
5.2.1 Installation of CardioSoft
For CardioSoft start the Setup Program without parameter.
5.2.2 Installation of CardioSys
For CardioSys start the Setup Program with parameter "cardiosys" (for example setup
cardiosys). For Software Update start the Setup Program without parameter.
5.2.3 Installation of MicroLab
For MicroLab start the Setup Program with parameter "microlab" (for example setup
microlab). For Software Update start the Setup Program without parameter.
5.2.4 Update from 16 Bit CardioSys to 32 Bit CardioSys
With the following steps you can copy all patient data, procedure data and configuration
data from a 16 Bit CardioSys (Version V1.0 to V3.X) to a 32 Bit CardioSys (Version
V4.1).
1. Copy the directory "C:\CARDIO" and all subdirectories to a Backup Media (this
Backup Media must be compatible with Windows NT 4.0). Restore this directory to
"D:\CARDIO".
2. Copy the CARDIO.INI file from the Windows directory to a floppy disk and copy it to
the Windows NT directory.
3. Copy the WIN.INI file from the Windows directory to a floppy. Open this file with Editor
(notepad.exe). Open the WIN.INI file in the Windows NT directory with the Editor
(notepad.exe). Copy the section "[CARDIO]" and the section "[EGM_RECORD]" from
WIN.INI of the floppy the WIN.INI file in the Windows NT directory. Delete the file
CA_SYS.INI in the Windows NT directory (if available).
4. Copy the file PERM_MEM.DAT from the Windows directory to a floppy and copy it to
the Windows NT directory.
5. Install CardioSys or MicroLab to the target directory "D:\CARDIO".
5.2.5 Installation of Communication Server
The Communication Server is a separate PC with up to 4 Modems connected. Rest ECG
procedures can be send from MicroSmart, CardioSmart, MAC 1100, MAC 1200 to this
Communication Server over a Modem. The Communication Server stores the Rest
ECG's in a local network server and can this data send to MUSE. This PC can only
used as Communication Server and cannot used for other purpose.
Prerequisites: CardioSoft V4.1, PC with minimum requirement see CardioSoft User
Manual, Up to 4 serial ports. Up to 4 Modems (important all Modems must be from the
same type), PC must be integrated in a local network to store the data or send the data
to MUSE.