Operating instructions
Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 23
Servicing Instructions 227 436 37 Rev. B V1.3
5 Installation
5.1 Installation of the hardware components
All connectors (including the patient lead) for system components and additional
instruments are located under the cover flap. Important: The cover flap may only be
opened by the technical staff. The system is supplied ready for operation. Only the
monitor, PC keyboard, function key keyboard (optional), mouse and any required
additional instruments need to be connected. To do this proceed as follows:
Open cover flap. Release lock with the appropriate tool.
* Monitor:
Place monitor in the monitor mounting and secure with the tilt-safe holder. Connect the
monitor power cord (emerging from the cable runway) to the monitor. Insert the monitor
signal cable into the cable runway and connect up to the monitor interface.
* Mouse
Place the mouse of the pull-out shelf. Insert the cable into the cable runway. Put
ferrite on the end of the cable. Connect cable to the mouse interface (PS2).
* Function key keyboard (optional)
Place function key keyboard on the pull-out shelf. Insert function key keyboard cable
into the cable runway and connect it up to the keyboard interface.
* PC keyboard (optionally with card reader)
Place PC keyboard on the pull-out shelf.
1.) Without function key keyboard:
Insert PC keyboard cable into the cable runway and connect up to the keyboard
interface.
2.) With function key keyboard:
Connect keyboard cable to the function key keyboard. Insert remaining keyboard
cable into the cable runway.
(Important! When using the PC keyboard with a card reader, set switch on
bottom of keyboard to "Mode 2".)
Important! The PC keyboard with a card reader must not be used in combination
with the function key keyboard.