3.4 SHIPPING PROTECTION
This ADK balance has protection for the weighing mechanism that must be
removed before the balance is used.
To remove the shipping
protection, remove the pan
then place the balance on
the table upside down.
Remove the cover over the
protection screw by
loosening 2 screws.
Remove the protection
screw, place in a safe place
in case it is needed for
shipping the balance again.
Replace the cover.
3.5 VOLTAGE SELECTION
This balance now includes a switch to select the mains voltage.
Verify the switch is set to the correct voltage for your area.
The switch is located on the back panel under a small cover.
Remove the cover on the rear of the balance.
Remove the cover by removing the small screws on the cover. Do
not remove the screws on either side of the cover.
Select either 110V or 220V as required. The 110V position is for
power supplies from 110 - 120VAC and the 220V position is for
voltages of 220 – 240VAC.
RS-232 connector 110/220V Select Fuse
1/2A
On/Off
Weigh below
Attachment
Shipping Protection
ADK BALANCE