standard repeat the calibration procedure.
If it still displays “CAL -no” send the balance to the service centre.
5.3 CHANGING CALIBRATION MASS VALUE
If you do not have the accurate calibration mass, you can change the
value of the calibration mass by using the following procedure. This will
enable to carry out the calibration of your balance accurately in
absence of the proper calibration mass.
Procedure
Turn the balance on by pressing [On/off] with the pan empty.
Press the [Cal] key. “CAL - -0” will be displayed.
Press [Cal] again. “20000.0” will be displayed.
Use the [Mode] key to increase the value up to 20001.5 or use the
[Print] key to decrease the value to 19998.5. The value will
change by one digit every time these keys are pressed during
setting up.
When the desired value is displayed (for example, 20000.8) press
[Tare]. “CAL - -0” will be displayed.
Press [Tare] again. “CAL F” will be displayed.
Place the calibration weight weighing 20000.8 and press [Tare].
First “CAL Endwill be displayed and after some time “20000.8” will
appear.
Remove the weight from the pan. It will display “0.0”.
Balance returns to the weighing mode.
Note: The value “20000.8” will not be recalled next time you calibrate
the balance.