Product Manual

6
©Adam Equipment 2017
The user can print the current configuration by pressing the [Print] key during the process providing
the second function for that key is not needed.
STOP TIME
Min 30
STAB RANGE stAB time
02
sec15
stop time
stop RANGE stop time
min 30
02 sec15
Display, shows StOP
Options are: Time, Stab or Time /
Stab
Option TIME
Will stop at a set time
Use arrow keys,
[Dspl] and [Test],
to change up and
down. The [Print]
and [PST] key to
move to the right
or left. Press
[Setup/Enter] to
continue.
Option STAB
Stops automatically when a stable
reading is recorded for a set time.
User can define the time and
stabilit
y
ran
g
e.
Option TIME / STAB
A combination of the other two
settings, a max time and
stability setting.
Display, shows Start
Options are: Manual (when the
start button is pressed), or
automatically (once the cover is
lowered after a 5 second delay the
test will start).
Display, shows PRINT TEST
Options are: on, off
Display, shows FORMAT
Options are: comp, summ
Note: PRINT TEST allows for the printing / saving the
results during the test. If set to on the PMB will also store the
results to memory using the file name given by the user. See
section on USING THE INTERNAL MEMORY SETTING.
The FORMAT option lets the user select the type of printing
output. Comp shows all the results and Summ shows just
the final results. See the complete manual for information on
the different output settings.