2.3 RUNNING A TEST
Once the test procedure has been set then a test can be run using the current settings. A test can
also be recalled from memory See USING THE INTERNAL MEMORY.
Place an empty sample pan on the analyzer.
Press [Tare] to zero the weight.
Remove and fill the sample pan with the sample.
Place the filled pan on the analyzer.
Lower the lid of the weighing chamber,
a. If automatic start has been selected the test will begin after a 5 second delay.
b. If Manual start has been selected press the [Start] key.
The display will show the current test settings if the Preview function is on, then the test will begin.
See section 5, Menu Structure.
Rapid Function
If the rapid heat function has been set to MANUAL the user can select if they wish to enable that
option now.
The display will show the last setting used, either:
Use the [Up] or [Down] keys to change the option, then press [Enter] to continue the test.
Alternate Displays
During the test the [Dspl] key can be pressed to see other information such as current mass, other
results such as %Solid, time remaining if time limit is set for the test. Each alternate display is shown
for 5 seconds then normal display is returned. If [Dspl] is pressed again the next alternate display is
shown.
The Alternate displays in order are:
Mass
% Moisture
% Solid
% Atro Moisture
% Atro Solid
Time Remaining for test if time limit set
Stopping the Test
During the test the user can stop the test at any time by pressing the [Start] key again. Otherwise it
continues until the stop condition is met.
MASS 23ºC
0.000g
MASS 23ºC
5.670g
OFF ON
rapid rapid