6.8 CALIbr = CALIBRATION ROUTINE
To set the parameter press the [Func] key to showCALIbr”.
Remove all weight from the platform.
Press [Print] and the display will show noCAL while the scale is
measuring the initial mass (pan must be empty).
Then the display will show LOAd xxx where xxx is the required
calibration weight.
Place the weight on the platform then press [Print] display will
show CAL” and then (after calibration) unLoAd”.
Remove the calibration weight the scale will return to normal
weighing.
6.9 rePL = RESULTS PRINTED AUTOMATICALLY OR MANUALLY
This parameter controls the operation of the RS-232 interface.
The balance can be configured to either print automatically when
the balance becomes stable or only when [Print] is pressed.
To set the parameter press the [Func] key to show “rePL”.
Press [Print] to see the parameter.
Press [Print] to select either “rePL= 0or “rePL= 1”.
rePL= 0 Manual output when [Print] key is pressed.
rePL= 1 Automatically print when the results are stable.
Press the [Func] key to return to weighing.