Product Manual
Table Of Contents
- 1.0 INTRODUCTION
- 2.0 TECHNICAL SPECIFICATIONS
- 3.0 UNPACKING THE SCALE
- 4.0 SETTING UP THE SCALE
- 5.0 KEYPAD AND DISPLAY
- Description
- Symbol
- 6.0 FUNCTIONS
- 6.1 bL = AUTOMATIC BACKLIGHT CONTROL
- 6.2 AVE = DISPLAY AVERAGING RATE
- 6.3 FIL = ADC FILTER RATE
- 6.4 bod = BAUD RATE SELECTION
- 6.5 cont = CONTINUOUS PRINTING
- 6.6 PIECE = PARTS COUNTING
- 6.7 StEPS = CHECK-WEIGHING FUNCTION (LOW-OK-HIGH)
- 6.8 CALIbr = CALIBRATION ROUTINE
- 6.9 rePL = RESULTS PRINTED AUTOMATICALLY OR MANUALLY
- 6.10 StAb = PRINT WHEN STABLE or INSTANTANEOUS
- 6.11 Aut = AUTOZERO FUNCTION
- 6.12 t1 = AUTOMATIC POWER SWITCH OFF
- 6.13 toP = HOLD MAXIMUM VALUE
- 6.14 nE = WEIGH IN NEWTONS
- 6.15 SuPP = Power supply voltage to PCB
- 7.0 COMMUNICATION WITH A COMPUTER/PRINTER
© Adam Equipment Company 2005
12
6.8 CALIbr = CALIBRATION ROUTINE
• To set the parameter press the [Func] key to show “CALIbr”.
• Remove all weight from the platform.
• Press [Print] and the display will show “noCAL” while the scale is
measuring the initial mass (pan must be empty).
• Then the display will show “LOAd xxx” where xxx is the required
calibration weight.
• Place the weight on the platform then press [Print] – display will
show “CAL” and then (after calibration) “unLoAd”.
• Remove the calibration weight – the scale will return to normal
weighing.
6.9 rePL = RESULTS PRINTED AUTOMATICALLY OR MANUALLY
• This parameter controls the operation of the RS-232 interface.
• The balance can be configured to either print automatically when
the balance becomes stable or only when [Print] is pressed.
• To set the parameter press the [Func] key to show “rePL”.
• Press [Print] to see the parameter.
• Press [Print] to select either “rePL= 0” or “rePL= 1”.
“rePL= 0” Manual output when [Print] key is pressed.
“rePL= 1” Automatically print when the results are stable.
• Press the [Func] key to return to weighing.