7.0 CALIBRATION
•
Calibration before use is recommended for best accuracy. It is suggested
that the scale be calibrated when first installed and any time the scale is
moved to a different environment.
• Ensure weights are ready before calibrating, the weight should be of a known
accuracy, for example OIML class M1 or Class 4 of ASTM E617.
• Do not calibrate the scale if a proper calibration weight is not available.
7.1 AUTOMATIC EXTERNAL CALIBRATION
In the weighing mode, press [FUNC/↑] several times until the display shows ‘CAL' to access
calibration mode.
• Press the [Tare/↵] key, The display will show “Pin”.
• Press the [FUNC/↑] key, [Tare/↵] key and [UNIT/→] key to enter calibration
‘Pin’
• The display will show “CAL 0” and when the stable symbol is displayed press
the [Tare/↵] key to confirm.
• The display will now show the calibration mass expected. ”CAL 01 kg”
(or “CAL 01 Lb”). Press the [FUNC/↑] key to change the calibration values if
necessary. Each model has 3 different kg and lb calibration selection values.
Calibration selectable values for all model:
LBX3 LBX6 LBX12 LBX30
1Kg/2kg/3kg 2Kg/4kg/6kg 5Kg/10kg/12kg 10Kg/20kg/30kg
•
Press the [Tare/↵] key to accept calibration mass, add the selected
calibration mass value as displayed to the pan.
• After the calibration mass is added, wait for the stable symbol to come on
and then press the [Tare/↵] key, the display will show “PASS”, and then return
to normal weighing after 3 seconds.