User Manual
© Adam Equipment Company 2018 7
• the scale will show “xxxxx.x” with the ‘HI’ LED lit. Otherwise the scale will
show “xxxxx.x” with the ‘OK’ LED lit.
• Press the [FUNC/↑] key to exit and go back to weighing mode.
5.3.4 HOLD FUNCTION
• When in weighing mode, press [FUNC/↑] until the display shows “X Hod”
• Press the [UNITS/→] key to go through the holding time of 5,10,20,30, AUTO
or Manual.
• If AUTO is selected the weight will be automatically held (locked on the
display) when stable until all weight is removed from the scale and the
[HOLD] key is pressed.
If a numeric value is selected the weight will be held only when the [HOLD]
key is pressed and it will be held for the number of seconds selected.
• For example, if you selected 5 the display will lock for 5 seconds after the
[HOLD] key is pressed. If you selected AUTO, the scale will lock when stable
and hold the value until you press the [HOLD] key.
• If Manual is selected the [HOLD] key will toggle the display to hold or release
the value when it is pressed.
• Press [FUNC/↑] to confirm the value and start hold weighing function.
6.0 FUNCTION SETTING
6.1 ENTERING THE FUNCTION SETTING MODE
•
Press [On/Off] to shut off the power.
• Press [FUNC/↑] after press [On/Off] to turn the scale on.
• The display will show “SLP 0”. Press the [Ta re /↵] key to go through the
backlight, g, lb, oz, lb:oz, bk au, pcs, per, chk.
6.2 SETTING OF AUTO SHUT OFF FUNCTION
• Press [FUNC/↑] to change the setting of this function then press [Tare/↵] to
confirm the change.
• “SLP 0” means that the function is disabled.
• “SLP X” means that the auto off function is on. The number is the time in
minutes before the scale will shut off automatically.