User Manual

© Adam Equipment Company 2018  10
7.0 CALIBRATION
Calibration before use is recommended for best accuracy. It is suggested
that the scale be calibrated when first installed and any time the scale is
moved to a different environment.
Ensure weights are ready before calibrating, the weight should be of a known
accuracy, for example OIML class M1 or Class 4 of ASTM E617.
Do not calibrate the scale if a proper calibration weight is not available.
7.1 AUTOMATIC EXTERNAL CALIBRATION
In the weighing mode, press [FUNC/] several times until the display shows ‘CAL' to access
calibration mode.
Press the [Tare/] key, The display will show “Pin”.
Press the [FUNC/] key, [Tare/] key and [UNIT/] key to enter calibration
‘Pin’
The display will show “CAL 0” and when the stable symbol is displayed press
the [Tare/] key to confirm.
The display will now show the calibration mass expected. ”CAL 01 kg”
(or “CAL 01 Lb”). Press the [FUNC/] key to change the calibration values if
necessary. Each model has 3 different kg and lb calibration selection values.
Calibration selectable values for all model:
LBX3 LBX6 LBX12 LBX30
1Kg/2kg/3kg 2Kg/4kg/6kg 5Kg/10kg/12kg 10Kg/20kg/30kg
Press the [Tare/] key to accept calibration mass, add the selected
calibration mass value as displayed to the pan.
After the calibration mass is added, wait for the stable symbol to come on
and then press the [Tare/] key, the display will show “PASS”, and then return
to normal weighing after 3 seconds.