Specifications
8 | P a g e © Adam Equipment Company 2008
7.0 TROUBLE
SHOOTING
7.1
ERROR
MESSAGES
If an error message is shown, repeat the step that caused the message. If the error
message is still shown then contact your dealer for support.
ERROR
CODE
DESCRIPTION POSSIBLE CAUSES
O-Ld
Overload Remove weight from the pan. Make
sure the machine is reading zero
before weighing. Do not over load
the pan. If the problem recurs
calibrate. Possible damage to the
mechanics.
OUt2
Out of Zero reading Make sure the pan is fitted correctly.
Make sure nothing is touching the
pan when turning on the unit.
Calibrate the machine. Possible
damage to the mechanics.
UnSt
Unstable Make sure the balance is on a flat
surface and away from vibration.
Possible damage to the mechanics.
LO or
Low Battery indicator Change the batteries or use on the
power adapter.
ErrE
E-Prom error Software error turn off and restart.
Contact Adam Equipment or your
dealer for further assistance.
No Power when turning on Change the batteries or use on the
power adapter.
7.2 REPLACEMENT
PARTS
AND
ACCESSORIES
If you need to order any spare parts and accessories, contact your supplier or Adam
Equipment. A partial list of the more common items follows:
Part Number Description
9402 AC adapter
(Country Dependent)
9401 Plastic pan