When a parameter is entered by pressing [Tare], the displays will guide you through the
parameter selected and the options available.
The parameter and their functions are:
PC (continuous): For continuous printing.
Cmd (from device): For printing from a device.
Available options when selecting the ‘PC’ option (use the [Tare] key to select).
4800: For setting the baud rate.
Available options when selecting the ‘Cmd’ option (use the [Tare] key to select).
4800: For setting the baud rate.
5.0 CALIBRATION
OIML TYPE APPROVAL: For the CKT-M models, the calibration is locked either by a sealed
jumper on the underside of the scale, or by a calibration count on the display. If the seal is
broken or tampered with, the scale needs to be re-verified by an authorised certification
body and re-sealed, before it is used legally. Contact your local metrology standards office
for further assistance.
5.1 CKT CALIBRATION
You need to enter a secure menu by entering a passcode when requested.
Press [Tare] once, during the initial counting of the display after the power is
turned on.
The display will show “P” requesting for the passcode number.
The fixed passcode is “1000”
Press the [Tare] key
The display will show “u-CAL”
Press the [Tare] key and the display will show “noload” to request all weight be
removed from the platform.
Press the [Tare] key to set the zero point
The display will then show the calibration weight suggested in the display. If the
calibration weight is different from the value shown, Press [CE] to clear the current
value then enter the correct value as an integer value, it is not possible to have
fractions of a kilogram or pound. For Example:
20kg =