10.0 ERROR CODES
During the initial power-on testing or during operation, the scale may show an error
message. The meaning of the error messages is described below.
If an error message is shown, repeat the step that caused the message, turning the balance
on, carry out the calibration or other functions. If the error message is still shown contact
your dealer for further support.
ERROR CODE DESCRIPTION POSSIBLE CAUSES
Err 1 Time input error. Tried to set an illegal time, i.e. 26hours
Err 2 Date input error Tried to set an illegal date, i.e. 36
th
day
Tl.zl Stability error Zero at power on not stable
Err 4 Initial Zero is greater than
allowed (typically 4% of the
maximum capacity) when
power is turned on or when
the [Zero] key is pressed,
Weight is on the pan when turning the scale on.
Excessive weight on the pan when zeroing the
scale.
Improper calibration of the scale.
Damaged load cell.
Damaged Electronics.
Err 5 Zeroing error Repower the scale to set zero
Err 6 A/D count is not correct
when turning the scale on.
Platform is not installed.
Damaged Load cell.
Damaged Electronics.
Err 7 Stability error Can’t wait until stable
Err 9 Calibration error The user calibration is outside allowed tolerances
for zero
Err 10 Calibration error The user calibration is outside allowed tolerances
for calibration
Err 18 PLU error Current weight unit is inconsistent with PLU unit,
can’t read PLU
Err 19 Incorrect weight limits set Weight lower limit is bigger than upper limit
Err 20 PLU 140 PLU storage/ reading is more than 140
Err ADC ADC chip error System can’t find ADC chip
--OL-- Overload error Weight over range
--LO-- Underweight error -20 division from zero it is not allowed