Product Manual

©AdamEquipment2018 18
6.3.6 Automatic Accumulated Totals
The scale can be set to automatically accumulate totals when a weight is placed on
the scale. This eliminates the need to press the [M+] key to store values into the
memory. However the [M+] key is still active and can be pressed to store the values
immediately. In this case the values will not be stored when the scale returns to zero.
See the Section 9.0 on RS-232 Interface for details on how to enable Automatic
Accumulation.
7.0 CALIBRATION
The CCT scales are calibrated using metric or pound weights depending on the region and unit
in use before calibration.
You need to enter a secure menu by entering a passcode when requested.
Press [Tare] once, during the initial counting of the display after the power is turned on.
The “Count” display will show “P” requesting for the passcode number.
The fixed passcode is “1000”
Press the [Tare] key
The “Weight” display will show “u-CAL”
Press the [Tare] key and the “weight” display will show “noload” to request all weight
be removed from the platform.
Press the [Tare] key to set the zero point
The display will then show the calibration weight suggested in the Count” display. If the
calibration weight is different from the value shown, Press [CE] to clear the current value
then enter the correct value as an integer value, it is not possible to have fractions of a
kilogram or pound. For Example:
20kg =
Press [Tare] to accept the calibration value and the “Weight” display will now show
“Load”.
Place the calibration weight on the platform and allow the scale to stabilize as indicated by
the stable indicator.
Press [Tare] to Calibrate.
When calibration is done the scale will restart and return to normal weighing.
After calibration, the scale should be checked whether the calibration is correct. If
Necessary, repeat calibration.