Product Manual
©Adam Equipment Co. 2019 14
Press [Func/Set] and scroll through the list of functions using the ⭡ and ⭣
directional keys
Press [Tare/↵] to enter animal 2 weighing. The display will show “LOAD” and the
Animal/Dynamic weighing symbol . The scale is now ready to weigh an
unstable animal or sample on a weighing scale platform.
To use the Animal Weighing function it is necessary to set the amount of filtering
required for the item to be weighed. More active animals will require a higher level
of filtering to give a stable result. Press the [Disp/Mode]. The display will show “Flt
x” where x is a value from 0 to 5. The higher the value the greater the amount of
filter will be. To increment the value shown press the [⭡] key then press the
[Tare/↵] key to accept it.
6.11.1 Animal 2 weighing procedure
With the weighing scale platform empty the indicator display will show “LOAD”.
Place containers or blankets onto the platform and press the [0/Esc] key to
remove the weight of the containers or blankets, alternatively, a long press on the
[Tare/↵] key, will retain the container or blanket value as “NET”.
Place the animals or sample to be weighed on the platform.
Press [Tare/↵] key to start to the measurement. The display will show the live
measurements until a stable weight is determined. The time it takes for the stable
value will depend upon the setting of the filter parameter in the first step.
When a stable reading is found, the display will show this value, and the display will
be locked for 2 seconds. The display will show the “Hold” symbol while the display
is locked.
After the 2 seconds delay, the display will now show “UNLOAD”, Remove one of
the Animal or items, and press [Tare/↵] key to calculate the weight of the removed
animal or item. The display will show the live measurements again, until a stable
weight is determined.
When a stable reading is found, the display will show this value, and the display will
be locked for 2 seconds. The display will show the “Hold” symbol while the display
is locked.
After the 2 seconds delay, the scale will automatically print out the result. If all the
animals or items have been removed, the scale will restart the function and it will