③If you want to switch from normal mode to expert mode, you need to operate
on the APP. In an expert mode, the first and third mode indicators are lit at the same
time.
④When the power is changed from off to on, the mode defaults to the mode used
before the last shutdown.
The running time of each mode is 10 minutes by default and can be set on the APP
from 1 to 20 minutes. After set up, the operation on the device or APP will run
accordingly.
5.7 Status Detection: When the electrode split away from the skin during the device
is working, the indicator light will flash, and the device will automatically detect
whether the electrode piece is reattached to the skin. If it has been reattached to the
skin, it will enter the current usage mode otherwise it will shut down after 60
seconds automatically. In this state, the device cannot perform operations such as
pause, mode switching, or any current intensity changes.
5.8 Low Battery: When the power is lower than 10%, the device will automatically
stop the current output and all indicator lights will turn off. The device will shut down
after buzzer sounds "Di-di" for 10 seconds. When in a low-battery state, please
connect to a charger for charging.