10.3.9 Configurando Comunicação
INTFPRTCL
SET/GET
Changes or verifies which communication interface the reader should use. Values
[WIEGAND, WG, ABATRACK, ABA, GPO]. .
Note: None (GPO) to use w1/DATA and W0/CLK signals as digital outputs triggered via ASCII
messages.
Example:
Direction
Description
$ GET INTFPRTCL<CR>
Host →
Reader
Host sends the message to check
which interface it is using.
$ WIEGAND<CR>
Host ← Leitor
Reader responds.
$ SET
INTFPRTCL=WIEGAND<CR>
Host →
Reader
Host sends the message to change
the parameter.
$ OK<CR>
Host ← Reader
Reader responds with OK.
WGBILTEN
SET/GET
Changes or checks the number of bits of Wiegand communication. Values [26, 34].
Note: Parameter used only when the communication interface type is Wiegand.
Example:
Direction
Description
$ SET WGBILTEN=26<CR>
Host → Reader
Host sends the message to change the
number of bits to 26..
$ OK<CR>
Host ← Reader
Reader responds with OK, successful
change.
ABADIGLEN
SET/GET
Changes or checks the number of digits of the Abatrack interface (CLOCK/DATA).
Values [10, 14].
Note: Parameter used only when the communication interface type is Abatrack.
Example:
Direction
Description
$ SET ABADIGLEN=14<CR>
Host → Reader
Host sends the message to change the
number of digits to 14.
$ OK<CR>
Host ← Reader
Reader responds with OK, successful
change.
STRDT
SET/GET
Changes or checks the Data String that the reader sends over the Ethernet
interface at each data reading. Data String uses keywords that are replaced by read
results.
Example:
Direction
Description
$ SET STRDT=id do tag:IDWG<CR>
Host → Reader
Host sends the message to change the
data string.
$ OK<CR>
Host ← Reader
Reader responds with OK.