al a r m In d I C a T o r s a n d sC r e e n sy m b o l s
Visual, Audio
Indicators
Description What To Do
3 short
beeps
+
1 long beep
No Breath Alarm
This alarm occurs when a breath
is not detected for a period of 2
minutes (120 seconds) or more�
This alarm can be silenced by
pressing the Alarm symbol
(the bell symbol) shown on the
screen� This alarm becomes silent
as soon as a breath is detected�
If no breath is detected after
approximately 30 minutes, the
unit shuts down to conserve
power� (This alarm is inactive for
ten minutes after startup�)
Check the
connection from
the cannula to
the device� Make
sure the nasal
cannula is properly
positioned on
your face and that
you are breathing
through your
nose� Make sure
the cannula tubing
is not kinked or
obstructed�
3 long beeps
+
1 short beep
Oxygen Concentration Alarm
This alarm occurs when the
device is delivering a lower
concentration of oxygen than
specied� This alarm can be
silenced by pressing the Alarm
symbol (the bell symbol shown
on the screen)�
Change to another
source of oxygen
and contact
your equipment
provider�