RemRest™ Tiers 1-3 (Dealer Manual) 900 Series CAUTION: USA Federal law restricts this device to sale by or on the order of a physician. 22090 Rev.
Table of Contents Specifications 2 Tier 1 3 Tier 2 & 3 6 SPECIFICATIONS SIZE 4.7" H X 8.6" W X 9.4" D 11.9cm H x 21.8cm W x 23.9cm D POWER 110-240VAC, 50-60Hz, 12V DC input PRESSURE RANGE 3.
RemRest Set Up Tier 1: ON/OFF LIGHT RAMP LIGHT To enter the Technician Mode, plug one end of the AC power cord into the AC inlet plug located on the back of the RemRest device. Plug the other end into the wall socket. The green light will flash on for two seconds then will remain off.
TO SET THE PRESCRIPTION PRESSURE (BLINK ONCE/OFF 3 SECONDS): The default setting is 10cmH2O. Using a manometer and pushing the up and down arrows, adjust the pressure to the prescribed setting. TO SET THE RAMP STARTING PRESSURE (BLINK TWICE/OFF 3 SECONDS): The default setting for the starting ramp pressure is 3.5 cmH2O. To enter the starting ramp set mode, press the white dot once. The white dot is located between the on/off button and ramp button.
NOTE: Tier 1 model does not incorporate automatic altitude compensation. In order to insure that this device provides the expected pressures, the following adjustments to the pressure setting must be made to the device: For an increase in altitude, please multiply the prescription pressure setting by the correction factor found in the table. For a decrease in altitude, please divide the prescription pressure setting by the correction factor found in the table.
Tier 2&3: LIGHT Note: There are no differences between the Tier 2 and Tier 3 label other than color. Tier 2’s label is aqua and Tier 3’s label is blue.
To enter the Technician Mode, plug one end of the AC power cord into the AC inlet plug located on the back of the RemRest device. Plug the other end into the wall socket. The light will turn on for two seconds then will turn off. The LCD will also turn on completely (as shown above) for 2 seconds and will then display prescription pressure.
TO VIEW OR RESET THE USAGE TIME: The 4 USE and TIME icons will flash on the LCD and the usage time the device was in use by the patient will be displayed on the LCD. This value should be zeroed for new patients to remove factory test hours. Press the up and down arrows at the same time to reset the usage time. Once the desired setting is reached, press the mode button to save the setting. TO SET THE RAMP STARTING PRESSURE: The default setting for the starting ramp pressure is 3.5 cmH2O.
LED light will blink 3 seconds on and 1 second off and will not stop blinking until the “data code” is read. The blinking will prompt the user to contact the technician. Instruct the customer to read the data code as follows: To view the “data code”, which is the usage time in hours, push and hold the MODE button for three seconds. A code will show up on the display for five seconds and the green light will flash. No other icons will turn on.
VIEW OR SET THE DYNAMIC CALIBRATION: This mode allows you to set a resistance compensation factor determined by the type of interface device used. When using a nasal mask or a full face mask, press the up or down arrows to select “r nr ”. If using a nasal device such as the Nasal-Aire I or II or the Spiritus, press the up or down arrows to select “r r ”, the blower function will compensate for device restriction. Once the desired setting is reached, press the mode button to save the setting.
MDSS GmbH Schiffgraben 41 30175 Hannover, Germany EC REP