Problem Solving Through Innovation! PresSsion Gradient Sequential Compression Units User Manual for Models: Multi 3 (4331-110V) (4332-220V) Multi 6 (4333-110V) (4334-220V) ISO 13485 CERTIFIED www.ActiveForever.
TABLE OF CONTENTS PresSsion® Gradient Sequential - Multi 3 & Multi 6 FOREWORD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 ABOUT COMPRESSION THERAPY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4 Precautionary Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4 Indications and Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . .
FOREWORD PresSsion® Gradient Sequential - Multi 3 & Multi 6 This manual has been written for the owners and operators of the Multi 3 and Multi 6 PresSsion® Gradient Sequential units. It contains general instructions on operation, precautionary practices, maintenance and parts information. In order to maximize use, efficiency and the life of your unit, please read this manual thoroughly and become familiar with the controls as well as the accessories before operating the units.
ABOUT COMPRESSION THERAPY PRECAUTIONARY INSTRUCTIONS The precautionary instructions found in this section and throughout this manual are indicated by specific symbols. Understand these symbols and their definitions before operating this equipment. The definition of these symbols are as follows: ! ! ! =CAUTION- =WARNING- =DANGER- Text with a “CAUTION” indicator explains possible safety infractions that have the potential to cause minor to moderate injury or damage to equipment.
ABOUT COMPRESSION THERAPY CAUTION PresSsion® Gradient Sequential - Multi 3 & Multi 6 WARNING • Use of controls or adjustments or performance of procedures other than those specified herein may result in injury. • DO NOT permit any foreign materials or liquids to enter the units. Take care to prevent any foreign materials including, but not limited to, inflammables, water, and metallic objects from entering the units. These may cause unit damage, malfunction, electrical shock, fire, or personal injury.
ABOUT COMPRESSION THERAPY PresSsion® Gradient Sequential - Multi 3 & Multi 6 DANGER WARNING • DO NOT connect the units to an electrical supply without first verifying that the power supply is the correct voltage. Incorrect voltage may cause unit damage, malfunction, electrical shock, fire, or personal injury. The units were constructed to operate only on the electrical voltage specified on the Voltage Rating and Serial Number Plate.
SPECIFICATIONS PresSsion® Gradient Sequential - Multi 3 & Multi 6 Specification Power Source & Frequency Fuse Electrical Type Electrical Class Power Consumption Pressure Inflation Time Deflation Time Treatment Time Number of Pressure Lines Height* Width* Depth* Standard Weight Multi 3 Multi 6 110-120 VAC, 60 Hz (Model 4331) 110-120 VAC, 60 Hz (Model 4333) 220-250 VAC, 50 Hz (Model 4332) 220-250 VAC, 60 Hz (Model 4334) T 0.5 Amps T 0.
NOMENCLATURE PresSsion® Gradient Sequential - Multi 3 & Multi 6 9 10 1 8 6 2 3 4 5 6 7 Blanking Plug - Remove this plug in order to inflate or deflate a second garment. Keep this plug inserted when inflating or deflating only one garment. 2. Air Supply Connections - Point of attachment for air hoses. 3. Power Supply Cord Receptacle - Location of Power Input from Power Supply Cord. 4. Power Switch - ON/OFF toggle switch. 5. Pressure Regulator Dial - Sets pressure. 6.
OPERATION PresSsion® Gradient Sequential - Multi 3 & Multi 6 1. Place the unit on a firm and level surface. 2. Plug the power cord (female end) into the power cord receptacle. This is located on the left side of Multi 6, and on the front of the Multi 3. 3. Plug the other end of the power cord (male end) into a grounded, electrical outlet. NOTE: The pump operates best at room temperature. If the unit has been stored below normal temperature, allow it to warm to room temperature for proper operation.
OPERATION PresSsion® Gradient Sequential - Multi 3 & Multi 6 7. Place the patient’s limb into the garment and close the garment with the zipper. The full leg or full arm garment should be worn as high up to the groin or armpit as possible, avoiding bunching or wrinkling of the garment in these areas. Any extra length at the foot or hand will fill with air and compress properly. NOTE: For comfort, it is recommended that an absorbent stocking, sleeve, or tubular bandage be worn beneath the garment. 8.
OPERATION PresSsion® Gradient Sequential - Multi 3 & Multi 6 11. Using the Therapy Timer Control dial (on the Multi 6 model), set the prescribed time that the patient is to receive compression therapy. 12. Press the Power Switch to the ON position. This starts compression therapy on the Multi 3 model. On the front of the Multi 6, the buzzer will sound. You must press the Start button to begin therapy.
OPERATION PresSsion® Gradient Sequential - Multi 3 & Multi 6 15. When the treatment is completed (the buzzer will sound, the pressure display light will turn off, and the garment will automatically deflate on the Multi 6 model), press the Power Switch to the OFF position. This will automatically deflate the garment. NOTE: Leave the pressure control at its position to use the same pressure for the next treatment.
TROUBLESHOOTING & MAINTENANCE PresSsion® Gradient Sequential - Multi 3 & Multi 6 CAUTION Electrical shock hazard. Removal of cover and servicing of internal components must be done by qualified service personnel. Unplug the unit before attempting to troubleshoot. Problem Will not operate. Pump operates, but garments will not inflate Possible Cause Unit not properly plugged in. Possible Solution Check plug. Insert firmly into receptacle. Check nameplate for proper operating voltage. Return to dealer.
ACCESSORIES Uniform Garments (for use with the Multi 3 only) 43013-Half Arm-17 in (43 cm) 43014-Full Arm-29 in (74 cm) 43016-Foot and Ankle-10 in (25 cm) 43017-Half Leg-20 in (51 cm) 43018-3/4 Leg-26 in (66 cm) 43019-Full Leg-29 in (74 cm) 43020-Full Leg-32 in (81 cm) 43021-Full Leg-35 in (89 cm) 43023-Full Leg-Wide-35 in (89 cm) PresSsion® Gradient Sequential - Multi 3 & Multi 6 Sequential Garments 43032-Full Leg 6 Chambers-32 in (81 cm) 43034-Half Arm 3 Chambers-17 in (43 cm) 43035-Full Arm 6 Chambers-
WARRANTY PresSsion® Gradient Sequential - Multi 3 & Multi 6 Chattanooga Group ("Company") warrants that the PresSsion units are free of defects in material and workmanship. This warranty shall remain in effect for one year (12 months) from the date of original consumer purchase.
Bringing Innovation to the Rehab World! ™ ISO 13485 CERTIFIED 4717 Adams Road P.O. Box 489 Hixson, TN 37343 U.S.A. 1-800-592-7329 Tel (423) 875-5497 FAX 73661B © 2005 Encore Medical, L.P.