User guide
Table Of Contents
- Special Notes
- Register Your Product
- Typical Product Parameters
- Label Location
- Section 1— General Guidelines
- Section 2— Operation
- Installing the Mattress Replacement System
- Installing the Side Rails
- Installing the Power Unit
- Connecting the Hose
- Connecting the Power Cord
- Using the Front Panel
- Power Button
- Turn Button (MA90Z Power Unit)
- Turn Time Button (MA90Z Power Unit)
- Turn Angle Button (MA90Z Power Unit)
- Rotation Button (MA95Z Power Unit)
- Select Button (MA95Z Power Unit)
- Hold Button (MA95Z Power Unit)
- Firm/Soft Buttons
- Automatic Wireless Fowler
- Mode Button (MA90Z Power Unit)
- Max Inflate Button (MA95Z Power Unit)
- Lock/Alarm Silence Button
- Power Fail LED
- Low Pressure LED
- Displaying the Total Run Time (MA95Z Power Unit)
- Powering Up the System
- Placing the Patient on the Mattress
- Transferring Patient From/To a Gurney
- Transferring Patient From/To a Wheelchair
- Preparing for CPR Procedure
- Section 3— Maintenance and Troubleshooting
- LIMITED WARRANTY

SECTION 2—OPERATION
MA90 and MA95 Series Mattresses 20 Part No 1148139
Placing the Patient on the Mattress
1. PresstheModebuttonorMaxInflatebuttontoturnonthecontrolunittomaximum
flow.
NOTE:Inthismode,theMaxInflateLEDlightsup.
2. Afterthemattressisfullyinflated,placethepatientonthebed.
3. Ensurethatthepatient’sfeetaretowardsthefootendofthemattress(theendwiththe
hose).
4. Centerthepatientonthebedfromside‐to‐sideandhead‐to‐foot.
NOTE:Specialpositioningmayberequiredwithcontractedpatientstoprovidecomfortable
positions.
5. Afterplacingthepatient,makecertainnoobjectswillfallunderthepatient,suchas
feedingtubes,IVʹsetc.
6. Waitfiveminutesforthemattresspressuretostabilize.
7. Setthecomfortpressuretothedesiredcomfortlevel.
8. Verifythatthepatienthasnotbottomedoutbyperformingthefollowingsteps:
A. Ensurethatthepatientislyingflatonhis/herbackinthemiddleofthemattress.
B. Placefourfingersbetweentheaircushionsdirectlyunderneaththesacralregion
ofthepatient’sbody.
C. Ensurethatthereis3to4‐fingerwidthclearancebetweenthebottomofthepatient
andthebedframe.
D. Adjustthecomfortsetting,ifneeded.
E. Waitfiveminutesforthemattresspressuretostabilize.
F. RepeatSTEPSA‐Euntilpatienthasnotbottomedoutandpatientcomfortis
achieved.
9. Ifthepatientfeelsthatthebedistoosoft/hard,presstheSoftorFirmbuttontoadjust
thecomfortsettings.
10. Usearegularpillowtohelpsupportandstabilizethepatientʹshead.
Transferring Patient From/To a Gurney
WARNING
ALWAYS engage the wheel locks of the bed and the wheel locks of the gurney
before transferring the patient between the bed and the gurney.
1. Engagethewheellocksofthebed.Refertotheowner’smanualprovidedwiththebed.
2. Engagethewheellocksofthegurney.Refertotheowner’smanualprovidedwiththe
gurney.
3. PresstheMaxInflatebuttontoachievemaximummattresspressure.