User guide
Table Of Contents
- Special Notes
- Register Your Product
- Typical Product Parameters
- Label Location
- Section 1— General Guidelines
- Section 2— Operation
- Installing the Mattress Replacement System
- Installing the Side Rails
- Installing the Power Unit
- Connecting the Hose
- Connecting the Power Cord
- Using the Front Panel
- Power Button
- Turn Button (MA90Z Power Unit)
- Turn Time Button (MA90Z Power Unit)
- Turn Angle Button (MA90Z Power Unit)
- Rotation Button (MA95Z Power Unit)
- Select Button (MA95Z Power Unit)
- Hold Button (MA95Z Power Unit)
- Firm/Soft Buttons
- Automatic Wireless Fowler
- Mode Button (MA90Z Power Unit)
- Max Inflate Button (MA95Z Power Unit)
- Lock/Alarm Silence Button
- Power Fail LED
- Low Pressure LED
- Displaying the Total Run Time (MA95Z Power Unit)
- Powering Up the System
- Placing the Patient on the Mattress
- Transferring Patient From/To a Gurney
- Transferring Patient From/To a Wheelchair
- Preparing for CPR Procedure
- Section 3— Maintenance and Troubleshooting
- LIMITED WARRANTY

SECTION 2—OPERATION
MA90 and MA95 Series Mattresses 18 Part No 1148139
Mode Button (MA90Z Power Unit)
1. PresstheModebutton()toselecttheMaxInflateModeortheLowAirLossmode.
•MaxInflateMode‐Inthismode,themattressinflatesrapidlytomaximum
firmness(pressurizedto35±6mmHg).Aseriesofbeepssoundeverythree
minutesasareminderthattheMaxInflatemodeisactive.After30minutes,the
MaxInflatemodedeactivates andthecontrolunitdefaultstotheprevioussetting.
NOTE:Ittakes5‐10minutesfortheMA90Zmattresstoinflatefully(inflationtimedepends
onsizeofmattress).
NOTE:MaxInflatemodecanbemanuallydisengagedbypressingtheMaxInflatebutton.
ThiswilldeactivatetheMaxInflateLED.
NOTE:ItisrecommendedthatMaxInflatesettingbeusedduringpatientingress/egress,
patientwoundcare,patientturningorpatientcleaning.
•LowAirLossMode‐Inthismode,themattressgoesintoon‐demandlowairloss
reliefmode.
Max Inflate Button (MA95Z Power Unit)
1. PresstheMaxInflatebutton()toselectthemaxinflatemodeandinflatethe
mattressrapidlytomaximumfirmness(pressurizedto35±6mmHg).Aseriesof
beepssoundeverythreeminutesasareminderthattheMaxInflatemodeisactive.
After15minutes,theMaxInflatemodedeactivatesandthecontrolunitdefaultstothe
previoussetting.
NOTE:Ittakes25‐60secondsfortheMA95Zmattresstoinflatefully(inflationtimedependson
sizeofmattress).
NOTE:ItisrecommendedthatMaxInflatesettingbeusedduringpatientingress/egress,patient
woundcare,patientturningorpatientcleaning.
2. Performoneofthefollowing:
• MA90ZPowerUnit‐PresstheMaxInflatebuttonagaintomanuallydisengage
MaxInflatemode.
• MA95ZPowerUnit‐SelectadifferentsettingtoexitMaxInflatemode,orpressthe
MaxInflatebuttonagain.
NOTE:ThiswilldeactivatetheMaxInflateLED.