Instruction Manual

33
Cleaning, Care and Maintenance Chapter 6
English
4. Install a replacement fuse, Inogen RP#125 (BUSS MDA-12) and reassemble the tip.
Ensure the retainer ring is properly seated and tightened.
Inogen One® G3 Column Change Procedure
NOTE Column change instructions are only to be used when maintenance is
required and are not intended for practice purpose.
1. Turn off the Inogen One® G3 concentrator by pressing the power button to shut
down the device.
2. Remove the Inogen One® G3 concentrator from the carry bag.
3. Remove the battery from the Inogen One® G3 concentrator.
4. Turn Inogen One® G3 over so that is standing in the
upside down position where the display panel is
facing down and the bottom of the Inogen One® G3
is facing up.
5. There are two columns (metal tubes) on the
Inogen One® G3 concentrator and one is located on
each side of the device.
Fuse
Cigarette Adapter Plug
Retainer Tip
Retainer
CAUTION For continued protection against risk of fire, only use fuse specified.
www.ActiveForever.com