Specifications

6
PRODUCT SPECIFICATIONS
Multi-Channel, Configurable, 12-Bit Data Loggers
FEATURES
Multiple channels
Monitor a variety of
parameters:
- Temperature
- Relative Humidity
- Current
- Voltage
- Pressure
- Process Signals
- Pulse
12-bit resolution
10-year battery life
3-year warranty
Up to 1.5 MB
memory
Fast sample rates
(up to 25 samples
per second)
Remote access via
modem or
Ethernet
Network capable
(connect up to 10
data loggers)
Alarm dial-out
User configurable
Front Back
Removable screw-type
terminal block for easy
connections
Rugged Noryl®
enclosure
Unique serial number
Input jack for interface cable
Internal thermistor
temperature sensor
Locking eyelet
Magnetic
backing
Size:
107 mm x 74 mm x 22 mm (4.2” x 2.9” x 0.9”)
Weight:
110 g (3.75 oz)
Case Material:
Noryl® plastic
External Connector:
Removable screw-type terminal strip with Common (-)
connection
Mounting:
Magnetic backing or locking eyelet
Operating Limits:
-40 to 70ºC (-40 to 158ºF) and 0 to 95% relative humidity
(non-condensing)
Clock Accuracy:
+/- 2 seconds per day
Battery:
3.6 volt Lithium, 1 Amp-Hour
Battery Life:
10 years under normal use (factory replaceable)
Power Consumption:
5 to 10 microamps (continuous)
Memory Sizes*:
32 KB (21,500 readings)
128 KB (87,000 readings)
1.5 MB (1,048,000 readings)
*See specific models for availability
Sampling Methods:
1. Continuous (First-in, First-out – not available with
sample rates faster than 8 seconds)
2. Stop when full (Fill-then-stop)
Sampling Rates:
User selectable rates from 25 per second (0.04 seconds) to
once every 8 hrs*
*BP-101 Battery Pack or PS-201 Power Supply required for
sample rates faster than 8 seconds (see Accessories section for
details)
Resolution:
12-bit (1 part in 4,096)
PC Requirements:
IBM PC or 100% compatible running Windows® 3.1,
’95, ’98, 2000, Me, NT or XP with at least 2 MB RAM,
2 MB of hard drive disk space and one free serial port
Software Requirements:
TrendReader® Standard (see page 34 for details)
Certifications and Standards:
- Certified to CE standard EN50082-1: 1997 (European
Generic Immunity) covering ESD, RFI, EFT/B, Surge,
Conducted Immunity, Power Frequency Magnetic Field
Immunity, and Voltage Dips and Interruptions
- Meets CE standards EN55011 Class B: 1997 (European
Radiated Emissions) and CISPR Class B: 1991
(International Radiated Emissions)
Resistance to X-Rays:
- Tested for protection against a 160kV dose @ 5mA for
30 seconds (150mA-sec) @ 38 inches F.F.D. (about 100
times that of an airport x-ray machine)
- Tested for protection against Gamma Ray (equivalent to
0.137-1.38 mega volts) IR 192 - 28 curies @ 30 second
exposure source to object distance 5”
INTERNAL TEMPERATURE SENSOR
(common on all models, except SmartReader Plus 9)
Type:
Negative Temperature Coefficient Thermistor
10,000 ohms @ 25ºC (77ºF)
Range:
-40 to 70ºC (-40 to 158ºF)
Accuracy:
+/- 0.2ºC over the range of 0 to 70ºC (+/- 0.3ºF over the
range of 32 to 158ºF)
Resolution:
0.03ºC at 25°C (0.05ºF at 77°F); better than 0.07ºC (0.12°F)
between -25 and 70°C (-13 and 158°F); better than 0.13ºC
(0.23°F) between -40 and -25°C (-40 and -13°F)
SmartReader Plus data loggers share many common features. For further specifications on individual models, refer to pages 7-14.
SmartReader Plus - Common Specifications