Specifications
11
SmartReader Plus 5
SmartReader Plus 6
FEATURES
• Three channels
• Monitor extremely
high temperatures
• Monitor temperature
over a wide range
• Use with a variety of
different thermo-
couple probes
• Greater resolution
and accuracy than
SR5
FEATURES
• Eight channels
• Record temperature
in eight different
locations
• Monitor extremely
high temperatures
• Monitor temperature
over a wide range
• Use with a variety of
different thermo-
couple probes
• Greater accuracy
and resolution than
SR6
Three-Channel Temperature (Thermocouple) Data Logger
The SmartReader Plus 5 data logger is ideal for those applications that require
monitoring high temperatures or monitoring temperature over a wide range. The
logger has one internal channel for measuring ambient temperature and two exter-
nal channels for use with ACR thermocouple wire or any J, K, S, T or E type
thermocouple wire. At 12-bit resolution, the SmartReader Plus 5 offers greater
accuracy and more precise measurements than the SmartReader 5.
PRODUCT SPECIFICATIONS
Eight-Channel Temperature (Thermocouple) Data Logger
Monitor temperature over a wide range with the SmartReader Plus 6 data logger.
Easily and accurately record temperature from any J, K, S, T or E type thermocou-
ple probe. At 12-bit resolution, the SmartReader Plus 6 offers greater accuracy and
more precise measurements than the SmartReader 6. With seven inputs for ther-
mocouple probes and an on-board cold-junction thermistor, this logger is one of
the most versatile thermocouple data loggers available.
No. of Channels:
Three - One for ambient temperature (also for cold-
junction compensation)
- Two for temperature (with thermocouples)
Accuracy:
+/- 0.5% of range + resolution*
Memory Size:
32 KB and 128 KB
Common Specifications:
See page 6
Software Specifications:
See page 34
Remote & Network Communications:
See page 38
Accessories:
ACR Thermocouple Wire or any J, K, S, T, and E
type thermocouple wire
Order Information
MODEL MEMORY CATALOG #
SRP-005 32 KB 01-0012
SRP-005-128K 128 KB 01-0121
Order Information
MODEL MEMORY CATALOG #
SRP-006 32 KB 01-0013
SRP-006-128K 128 KB 01-0122
SRP-006-1.5M 1.5 MB 01-0151
PRODUCT SPECIFICATIONS
No. of Channels:
Eight - One for ambient temperature (also for cold-
junction compensation)
- Seven for temperature (with thermocouples)
Accuracy:
+/- 0.5% of range + resolution*
Memory Size:
32 KB, 128 KB and 1.5 MB
Common Specifications:
See page 6
Software Specifications:
See page 34
Remote & Network Communications:
See page 38
Accessories:
ACR Thermocouple Wire or any J, K, S, T, and E type
thermocouple wire
THERMOCOUPLE TEMPERATURE (For both SmartReader Plus 5 and 6 )
T
ype Range *Resolution
J -50 to 600ºC (-55 to 1100ºF) 0.30ºC (0.55ºF)
K -100 to 1150ºC (-145 to 2100ºF) 0.50ºC (0.90ºF)
S 0 to 1450ºC (32 to 2642ºF ) 1.0ºC (1.8ºF)
T -200 to 400ºC (-325 to 750ºF) 0.40ºC (0.75ºF)
E -50 to 680ºC (-58 to 1250ºF) 0.24ºC (0.44ºF)