Specifications
SmartReader Plus Data Logger
66
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
For the internal channel, use equation 45
For the 100 ohm probes, use equation 86
For the 1000 ohm probes, use equation 87
For results in resistance (ohms):
For the internal channel, use equation 45
For the 100 ohm probes, use equation 84
For the 1000 ohm probes, use equation 85
NOTE: Even equation numbers are for 100 W probes and odd equation
numbers are for 1000 W probes.
4. Adjust the remaining setup options as desired then save the setup.
1.13.3 Specification of RTD Channels
The following specifications are for:
4-Wire Platinum 100 Ohm @ 0° C & TCR = 0.00385
4-Wire Platinum 1000 Ohm @ 0° C & TCR = 0.00385
Description: Value:
+SIG Input Impedance: >1 Meg. ohms
-SIG Input Impedance: >1 Meg. ohms
Excitation Current <=1mA
Temp. Range
(100 & 1000 W probes): -230 to 300°C (-382 to 572°F)
Accuracy ** (100 & 1000 W): ± 0.3°C (0.54°F) + resolution or
± 0.06 % full scale + resolution
Resolution (100 W & 1000 W): ± 0.14°C (±0.25°F)
Resistance Range (100 W): 5 to 210 W
Accuracy ** (100 W): ± 0.13 W + resolution or
± 0.06 % full scale + resolution
Resistance Range (1000 W): 50 to 2100 W
Accuracy ** (1000 W): ± 1.3 W + resolution or
± 0.06 % full scale + resolution
Resolution (100 W): ± 0.05 W
Resolution (1000 W): ± 0.5 W
* For internal channel specifications, please see Common Specifications for
SmartReader Plus loggers
** Temperature compensated for the range of -20 to 80°C (-4 to 176°F)