Specifications
SmartReader Plus Data Loggers
63
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
instead of the logger supplying the voltage potential, the logger detects the voltage level of the
pulse. The pulse must have a low level voltage less than 0.5 volts DC and the high level
voltage must be from 4.5 to 24 volts DC.
The minimum pulse length and minimum interval between pulses must be 2 milliseconds for
the slow input and 10 microseconds for the fast input. For the SmartReader Plus 9 to register
a count, the input voltage must go from a low level to a high level and back again to a low
level.
1.12.3 Setup and Use
The following steps will guide you through the procedure of setting up and using your
SmartReader Plus 9 Pulse Logger.
1. Always make sure that the circuit you want to monitor is compatible to use with the logger
before connecting it. Refer to the Specifications section for details on the type of
switches and logic signals that will work. Key things to note are:
· If you are going to be monitoring the opening and closing of switches, there must not be
a power source connected to the contacts (the contacts must be "dry").
· If you wish to monitor a circuit that produces pulsed signals, then the low level must fall
between 0 and 0.5 volts DC and the high level must be between 4.5 volts and 24 volts
DC.
· The frequency of the pulses or contact closures that you want to monitor must not
exceed the maximum capability of the SmartReader Plus 9 which is 4095 over any
sampling period.
2. Set the mode of operation of your SmartReader Plus 9 in TrendReader.
Channel
Description
Equation
0
Pulse Input 1
81
1
Pulse Input 2
81
3. Connect your switches or signals to check the operation of the logger by viewing the
Realtime reading while generating pulses. If the value increments, then you are ready to
start logging in the field.
1.12.4 Specifications
Pulse Inputs
Range: 4095 pulses/sample period
Accuracy: ±1 pulse/sample period
Absolute Max. Voltage: +24/-22 volts DC
Input Impedance: >1.0M for Vin<5V;