Specifications

SmartReader Plus Data Loggers
43
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
TrendReader guide for more information.)
5. Save the new setup.
1.7.3 SRP4 LPD Pressure Specifications
Type: Silicon piezoresistive strain gauge
Max. Pressure: Four times full scale
Media: Non-corrosive gases and dry air
Temp. Range:
Operating: -40 to 70°C (-40 to 158°F)
Compensated: 0 to 70°C (32 to 158°F)
Thermals Accuracy: ±0.5% FS (over compensated range)
Nonlinearity: ±0.05% FS (best fit straight line)
Hysteresis and Repeatability: ±0.05% FS (typical)
Long Term Stability: ±0.5% FS
Pressure Ranges and their Equations
Range in W.C.
Equation
Accuracy
-0.5 to +0.5"
13
± 0.05"
-1 to +1"
13
± 0.05"
-2 to +2"
88
± 0.05"
-5 to +5"
89
± 0.1"
-10 to +10"
37
± 0.1"