Service manual
LUMISCAN ACR-2000 SERVICE MANUAL – SECTION 6.0 PERIODIC MAINTENANCE
6-2
6.1.4 Clean area around the system daily. Check for dust.
6.2 Routine Maintenance
Maintenance Periods:
24 times every 12 months.
6.2.1 Cleaning Covers
The outside covers of the ACR-2000 READER should be cleaned with a mild soap or
detergent. Do not spray cleaner directly on the covers. Spray the soap or detergent on a
soft, clean cloth, then wipe down the covers.
6.2.2 Every other week, clean the insides of all the cassettes with a lint-free cloth to remove
dust and dirt.
6.3 Periodic Cleaning
Maintenance Periods:
4 times every 12 months.
WARNING
IT IS IMPORTANT THAT THE LUMISCAN COVERS
REMAIN ON THE SYSTEM AT ALL TIMES. THE COVERS
SHOULD ONLY BE REMOVED FOR SERVICE, AND THEN
IMMEDIATELY REPLACED. THIS WILL MINIMIZE DUST
ENTRY.
WARNING
THIS EQUIPMENT EMPLOYS A LASER. LASER
RADIATION MAY BE PRESENT IF THE LUMISCAN ACR-
2000 READER IS OPERATED WITHOUT COVERS.
AVOID LASER BEAM. DIRECT EYE EXPOSURE TO
LASER LIGHT MUST BE AVOIDED