LUMISCAN ACR-2000 READER SERVICE MANUAL - SECTION 5.0 SERVICE ADJUSTMENTS
5-1
____________________________________________________________________________
5.0 SERVICE ADJUSTMENTS
_________________________________________________________________
This section covers the LUMISCAN ACR-2000 READER components that can be adjusted in the field.
The following equipment is required for any alignment or adjustment performed in this section of the
service manual.
Requirements
HARDWARE REQUIRED
Computer with keyboard and monitor
Interconnect cable, PN 0062-052
Data Control PCA (DCB)
Oscilloscope, 50 Mhz, dual trace or better
Digital Volt Meter(DVM)
Imaging workstation
DVM
Laser power meter, Photodyne or equivalent
Storage phosphor image plates, 14 x 17inches, Agfa, Kodak or Fuji
Interlock switch cheat keys, 2 each
X-ray system capable of 85 kVp at 8 mAs
X-Ray exposure meter (capable of discerning 0.05 milliRads)
Copper step wedge, .020x 8 steps
Copper plate (1mm thick)
Set of Standard Screw Drivers( Flat blade, philips)
Small Screw Driver(Tweaker)
Hex Driver Set (5/16 thru .050)
Photographic Dark Cloth Cover
Diagnostics
Galvo Linearity Pattern P/N 0070-966
SOFTWARE REQUIRED
Lumisys diagnostic and calibration tools must be available. These are typically found in the
C:\LSDT32\TOOLS directory. The Data Acquisition (DACQ) board must also have its associated
CLUT file available, as generated during board-level testing.