Service manual
LUMISCAN ACR-2000 SERVICE MANUAL - SECTION 2.0 INSTALLATION
2-4
2.2.4 Media Support Assembly
Fasten the Media Support Assembly to the top of the Front Enclosure.
2.2.5 Screen Exit Assembly
Place the Screen Exit Tray in the brackets on the bottom of the Front Enclosure.
2.2.6 Installation of Data Control Board (DCB)
The ACR-2000 Reader is controlled by a proprietary Lumisys Data Control Board in an IBM type PC.
The DCB is 8 bit ISA. The DCB uses three resources, IRQ, Mem Seg Addr, and Base I/O Addr.
Note1: The ACR-2000 product contains a pre-configured PC workstation. It is not necessary to install the
Data Control Board unless a different PC is used.
Note2: If a replacement DCB is to be installed in a PC supplied by Lumisys, the PC will likely be set to
use certain resources for the DCB. These likely resources are IRQ 7, Mem Seg D800, and Base I/O 100.
It is recommended to set the DCB to these resources before installation. Please consult Appendix A for
the resource chart.
With power off, remove cover from the target PC or compatible(See Note below). Taking appropriate
anti-static precautions, remove the Data Control Board from the accessory box and install it in an 8 or 16
bit ISA slot. Secure the DCB in the slot and replace the covers.
The Data Control Board is factory configured to use IRQ 5 , I/O Addresses 100-11F and has a 32
KByte window at hex address D0000-D7FFF. If these addresses conflict with your system configuration
they may be changed. Please see APPENDIX A for instructions on how to change the IRQ level, the I/O
address, or the 32 Kbyte window address.
NOTE
The minimum hardware requirements for the Host
computer is as follows:
CPU: Pentium II 266MHZ
RAM: 128MB
Operating system: Windows NT 4.0 with Service Pack 4 or higher
Microsoft Internet Explorer 4.01 or higher
2.2.7 Installation of Control Interconnect Cable
Remove the Control Interconnect Cable (a 37 pin, Male/Male, D-Subminiature cable) from the
accessories box. Connect one end to the Data Control Board the other end to the LUMISCAN ACR-
2000 READER and secure both ends using a small flat blade screwdriver.