Service manual

LUMISCAN ACR-2000 SERVICE MANUAL - SECTION 2.0 INSTALLATION
2-2
LUMISCAN ACR-2000 READER.
With TWO PEOPLE, lift the LUMISCAN ACR-2000 READER off the pallet.
To install the LUMISCAN ACR-2000 READER, follow the Hardware Installation Procedures.
2.2 Hardware Installation
2.2.1 Tools Required
#1 flat-head screwdriver
#1 philips-head screwdriver
2.2.2 AC Voltage
You must verify that the LUMISCAN ACR-2000 READER is set up for the correct AC line voltage.
This can be checked by looking at power configuration panel which is located just to the right of the power
entry plug next to the on/off switch in the rear of the machine. The LUMISCAN ACR-2000 READER is
set for 120 VAC for domestic units. See Figure 2-1.
If the LUMISCAN ACR-2000 READER is to be operated at 200/240VAC line voltage configuration
requires removing and inverting the Corcom fuse module, and the programming card must be changed.
Note: disconnect power before changing line voltages or fuses. First remove the plastic cover/fuse
module by using a small screwdriver to pry out the fuse module(See figure 2-1). Loosen the philips head
screw and remove and invert the fuse block and tighten the screw. Two each 1 Amp 250V 5mm x 20mm
Slo-Blo fuses should all ready be installed, if not install them. Next remove the programming card rotating
it until the desired voltage is pointing inward and rotate the voltage indicator to point outward and reinsert
the card. Replace the fuse module and ensure the indicator is pointing to the correct line voltage. The
LUMISCAN ACR-2000 READER will is now set for operation.
Note
The fuse in the LUMISCAN ACR-2000 READER for 100/200 volt operation is a 1.5 Amp 250/75 volt
Slo-Blo fuse (Lumisys Part Number 0065-513). The fuses used for 200/240 volt operation are 1.0 Amp
Slo-Blo fuse (Lumisys Part Number 0068-487).
The LUMISCAN ACR-2000 ERASER comes either in a 120VAC model or a 240VAC model.
2.2.3 Power Cable
The LUMISCAN ACR-2000 READER utilizes an international IEC grade connector for the power
cable. Systems are shipped with a standard NEMA 5-15 hospital grade cable. This cable requires
replacement depending upon the country of installation. Insert the female end into the input power socket.