User's Manual

25
4.2.1.
Direct mounting
See Figure 25
Ensure a flat surface: remove paint layers. Step 1.
NOTE: if you use a spot facing tool be sure to not exceed 0.8mm depth [0.031 in].
Drill a hole (Ø5 mm and XXmm depth [Ø 0.19 in - ) Step 2.
Thread the hole with M6 tap Step 3.
Screw M6 stud with a length of XXmm Step 4.
CAUTION: The part of the stud inserted into the sensor can be up to 6 mm
[0.23 in]. A bottoming stud may cause base-strain and also potentially damage
electronics.
Screw the sensor on the stud and tighten at 5 - 7 Nm [44 - 62 inch-lb] using the Step 5.
dedicated tube wrench and your handgrip.
NOTE: for a better contact a film of grease/oil can be applied between sensor and
mounting surface (note the oil/grease must be compatible with the temperature of
the measurement point).
CAUTION: If you use a 44mm wrench [1.73 in] or an adjustable wrench be careful
to observe the indicated torque range.
Fill in the installation report with sensor information (S/N, type, location, Step 6.
orientation)
Figure 25: direct mounting with M6 stud
Figure 26: spotface mounting