User's Manual

23
4. SENSORS INSTALLATION
4.1. SENSOR LOCATION
The sensors are mounted on the equipment to monitor following the rules for vibration
sensor installation.
Leave enough space around the sensor to ensure the best transmission/reception possible
of radio signals. At least 100 mm around the device antenna (plastic cap). If the sensor has
to be installed in tiny spaces, the use of an expander might be a solution to improve the radio
coverage (if necessary).
Figure 23 : clear space around the sensor/expander
4.2. MOUNTING INFORMATION FOR SENSORS
A damp cloth should be used when installing or taking off the gateway in hazardous areas to
eliminate static electricity resulting from this operation.
Table 10 : mounting characteristics for sensors
Thread
M6-1, max depth 6mm [0.23 in]
Flat mounting surface
Ø32.5 mm x 1mm height [Ø1.28 in x 0.039 in]
Hex head
44 mm [1.73 in]
Fastening torque
5 7 Nm [44 62 inch-lb]
CAUTION:
Do not grasp the sensor by the plastic cap.
Do not hit the plastic cap.
Do not tighten the sensor by the plastic cap
5 mounting modes:
Direct mounting: M6 stud on a flat surface
Spotface mounting : for convex surface
Cementing pad: the easiest way to install
Triaxial mount: to position the triaxial sensor accurately.
Temporary mount: only for testing wireless coverage before final assembly.
Ø100 mm
clear space