User Manual

ACL 450B Operation Manual Rev. Jan 10, 2022 MKB
Page 3 of 6
OPERATION
1. Power ON:
Slide the power switch to the ON position and verify activity on the display. The kV/Inch light will
illuminate red when the meter is on, and the pushbutton is in the HOLD position.
With power ON, check the display. Battery low is indicated by all four decimal points illuminated.
If low battery is indicated, replace the battery before use.
2. Ground Meter
Grounding (discharging) your body also discharges the instrument through its dissipative case. If you
are not wearing a grounded wrist strap, discharge your body by touching a grounded metal object
such as a water pipe, conduit, or workbench. The meter can also be grounded by attaching a ground
cable to the snap on the back of the case and connecting the other end of the cable to a known
ground.
3. Zero adjust the instrument
Once grounded, with unit ON, point the sensor plate away from any charged object and press the
pushbutton so that it clicks into the lower MEASUREMENT position. Do not touch sensor, but
point it toward a known grounded surface at 1 inch until the concentric ring forms. Alternatively
point the sensor into free hand (this will shield the sensor from external fields). If the display does
not read 0.00, adjust the ZERO knob until the display reads 0.00. Once set, the zero should not
have to be reset unless the control has moved.