Instruction Manual

Automation Components, Inc.
2305 Pleasant View Road | Middleton, WI 53562
Phone: 1-888-967-5224 | Website: workaci.com
Page 2
Version: 6.0
I0000822
FIGURE 3: ENCLOSURE
MOUNTING
FIGURE 6: SENSOR MOUNTING
ON WALL
VAC
SENSOR MOUNTING (Continued)
Alternative, the senor may be mounted on walls using a 1/4” mounting clip (ACI Item #108169). The sensor
should be mounted in an area where air circulation is well-mixed and not blocked by obstructions. Slide
the sensor probe through the mounting clip - see FIGURE 6. Drill a 1/4” screw through the socket and
tighten to the wall. If a thermal buer is required, use a glycol kit (ACI Item #130127). Drill pilot holes for the
provided mounting screws. Use the mounting clip holes as a guide - see FIGURE 7. Drill the #10-16 x 1/2“
screws through the bracket holes and fasten it to the wall. Insert the bottle into the bracket, and make sure
it is seated securely.
Note: When using in tank or glycol application, the
sensor cannot be fully submerged. The end of the probe
must be kept above the liquid.
BULLET PROBE
1/4” MOUNTING CLIP
FIGURE 7: GLYCOL SENSOR
w/ BRACKET
SENSOR WIRES
GLYCOL BOTTLE AND
STANDARD 4” SENSOR PROBE
STEEL MOUNTING
BRACKET
FIGURE 4: WIRE ROUTING
FIGURE 5: SENSOR MOUNTING
WIRING INSTRUCTIONS
PRECAUTIONS
Transmitter must be powered by 24 VDC only.
Remove power before wiring. NEVER connect or
disconnect wiring with power applied.
When removing the shield from the sensor end,
make sure to properly trim the shield to prevent
any chance of shorting.
When using a shielded cable, ground the shield
ONLY at the controller end. Grounding both ends
can cause a ground loop.