Handbook Release 10.1 December 2007 WEST Carbon dioxide flux meter Continuous Monitoring ITA LY FOGO, FURNAS, TERCEIRA, MAMMOTH Horseshoe Lake, San MIGUEL, San VINCENTE, San SALVADOR, Santa ANA, MOUNUN , NYOS.
If your equipment requires maintenance in Italy: BEFORE YOU SHIP THE INSTRUMENT BACK TO ITALY REMEMBER: !When you ask your shipping agent to send the instrument to Italy check that on the Air Waybill the Airport of destination is PISA. Any other airport of destination creates a lot of problems in delivering the items (delay, costs, custom problems etc.) !Check that the Company your shipping agent chooses lands in Galileo Galilei Airport in Pisa.
INDEX INDEX Page List of the components Safety Information Hints 4 WEST Systems 5 6 Site selection Shelter and piquets kit Power supply Solar cell charge regulator Connections Installation of the Acc. Chamber Connection resume A.Chamber tubes 2.1 2.2 2.3 2.4 2.5-2.9 2.10 2.11 2.12 Pump Tubing / Fittings / Filters 3.1-3.2 3.3-3.4 Polytron IR CO2 Barometric pressure Pumping flow and filter status Soil Water content Soil temperature 4.1-4.2 4.3 4.4 4.5 4.
INDEX INDEX 7 Running FluxStation General considerations SD memory card StandBy and Running www.westsystems.com 8 Portable flux meter handbook Release 7.00 -FluxManager- July 2007 Calibration General considerations 8.1 Polytron IR CO2 gas calibration 8.2-8.6 Polytron IR CO2 4-20mA calibration 8.7-8.8 Flux meter calibration 8.10-8.14 9 Databases structure General considerations scada_log.mdb scada_data.mdb stations phisicalchannels channels telemetry 10 WEST Systems 7.1-7,2 7.3-7.4 7.5-7.
The package contents Shelter Feet Piquets Telemetry Backfire antenna Coax 3 meter N-N Coax 10 meter N-N SSR24 Slave SSR24 Master Quantity Check 1 1 1 1 WIND00075 1 Pluvio116 1 1 CS6161_21 1 1 C714 1 1 1 4 12 2427581 2 1 1 1 1 AC Accumulation chamber AC.Box 1 1 P.Supply Solar cell regulator 12V 40A/h battery Solar cell 1 1 1 FluxStation FluxStation AC Board PowerBoard CPU Board PAGE Shelter S/n DRGEEX02 C1130006 www.westsystems.
WEST Systems www.westsystems.com PAGE 1.2 Safety information ! ! Carbon dioxide is a TOXIC GAS, Carbon dioxide is colorless, odorless, tastless and is heavier than air. Air concentration higher than 5000 ppm can cause dizziness, shortness of breath, rapid pulse. Higher concentrations of carbon dioxide can be lethal. Diffuse carbon dioxide fluxes are normally related to anomalous carbon dioxide air concentration.
About installation Before to start the installation of the FluxStation is better to install the Master center telemetry systems and to setup the Scada software suite on the computer. 1.1 PAGE Hints The onfield installation is described on chapter 2.
WEST Systems www.westsystems.com PAGE 1.
On-Field Installation Continuous monitoring device handbook www.westsystems.com Release 10.1 November 2007 Select the installation site after a carbon dioxide flux survey. The best location is in the middle of an anomalous efflux area with an homogeneous flux distribution.The optimal flux value is in the range from 2 to250 moles per square meter per day.
On-Field Installation In the drawing the top view of the protection box (DHW), the sensor are not showed. To place the piquets: Unscrew the bolts and lift at the maximum elevation the shelter: After the installation of piquets down to the shelter at 10 cm distance from soil. lift Bolts Use the piquet-prolongue to hammer the piquets Soil Hit hardly here to block the 3 piquets WEST Systems WEST Systems Continuous monitoring device handbook www.westsystems.com Release 10.
2.3 Continuous monitoring device handbook www.westsystems.com Release 10.1 November 2007 In the picture one of the four stand with the 3 piquets. The soil on the picture is not exactly a “soft” soil... In the picture the “chair” for the FluxStation case that must be mounted on the right part of the shelter to put the fluxstation in a accessible position. In the picture the accumulation chamber connection box, that must be mounted on the shelter close to the accumulation ch.
WEST Systems WEST Systems Continuous monitoring device handbook www.westsystems.com Release 10.1 November 2007 Continuous monitoring device handbook www.westsystems.com Release 10.1 November 2007 PAGE 2.
On-Field Installation 2.5 PAGE Amp. 6+T Male Amp. 6+T Female from Solar Cell Amp. 6+T Female Battery - Battery + Aux.2 Aux.1 CE WEST Systems RS485/1 RS485/2 RS485/3 Battery Power supply cable Pin Function Color 1 +12V Red 2 Gnd Black 3 n.c. T n.c. Solar cell cable Pin Function Color 1 n.c. 2 Gnd Black 3 +12V Red T n.c. CH 4/5 CH 2/3 www.westsystems.com FL-I007 WEST Systems Acc.Ch. Antenna Continuous monitoring device handbook www.westsystems.com Release 10.
PAGE 2.6 On-Field Installation Solar cell charge regulator + - + - + The LED light is: Green when the battery is full Yellow normal charge Red when the battery is empty Continuous monitoring device handbook www.westsystems.com Release 10.1 November 2007 10 Continuous monitoring device handbook www.westsystems.com Release 10.1 November 2007 The LED light is ON when the solar cell is exposed to sunlight 10 Ampere fuse (car type) If the battery voltage fall down to 11.
On-Field Installation 2.7 PAGE Connections A B Power A CH3 WEST RS232 CPU CPU RUN MMC-SD ASF THOMAS ASF THOMAS CH1 CH0 WEST Pump Brombach CH2 POWER POWER ON PULL AND SWITCH POWER OFF +12 +24 bp_fff007 ON All electrical and pneumatic connections to the flux station are on the left side panel. Acc.Ch. RS485/1 RS485/2 RS485/3 Battery Antenna Aux.2 Aux.1 CH 4/5 CH 2/3 www.westsystems.
On-Field Installation The external sensors connections: The FluxStation is supplied with several meterological and soil parameters sensors: The FluxStation is supplied with several meterological and soil parameters sensors: Some of them are managed by the AirBox: !Air temperature and relative humidity; !Wind speed and direction; !Rain fall gauge; and the others by the SoilBox: !Soil temperature; !Soil water content; SoilBox and AirBox , described in detail in the Appendix S and Appendix A, are connected
On-Field Installation 2.9 PAGE Meteorological detectors: AirBox To FluxStation RS485/1 connector, using a RS485 cable. To Rain Gauge See page A.2 Continuous monitoring device handbook www.westsystems.com Release 10.1 November 2007 Continuous monitoring device handbook www.westsystems.com Release 10.
PAGE 2.10 On-Field Installation Soil.box Continuous monitoring device handbook www.westsystems.com Release 10.1 November 2007 HUB-04 Soil temperature probe (Page 4.8) WEST Systems To FLuxStation RS485/2 connector, using a RS485 cable (see page 2.6) www.westsystems.com OUT IN-B IN-A Soil water content CS616 (Page 4.7) Pt100 TDR WEST Systems SoilBox (Appendix S) RS485 Continuous monitoring device handbook www.westsystems.com Release 10.1 November 2007 The Soil.
On-Field Installation 2.11 PAGE Accumulation chamber Acc.Ch. RS485/1 RS485/2 RS485/3 Battery Aux.2 CH 4/5 Aux.1 CE Cable AC2 CH 2/3 www.westsystems.com FL-I007 Cable AC1 To Aux 1 To Acc. Ch. CloseSw Fan Aux Linak OpenSw Continuous monitoring device handbook www.westsystems.com Release 10.1 November 2007 Continuous monitoring device handbook www.westsystems.com Release 10.
On-Field Installation 04 TEN VV Place the collar and fill again the hole , taking care off the sealing of the collar. VV NE -T 40 1_rf 2_rf br-05 1_rf 2_rf The accumulation chamber base must be at the same level of the soil surface. Fill the collar with soil leaving the minimum empty space to avoid carbon dioxide accumulation inside the collar. br-05 1_rf 2_rf WEST Systems br-05 Continuous monitoring device handbook www.westsystems.com Release 10.
On-Field Installation 2.13 PAGE Connections resume 10 from Solar Cell Amp. 6+T Female Battery - Battery + Acc.Ch. RS485/1 RS485/2 RS485/3 Battery Antenna Battery Aux.2 Aux.1 CE x1 Au To Acc. Ch. CloseSw Fan Aux Linak www.westsystems.com FL-I007 To RS485/2 To To Aux 1 CH 2/3 h. .C c Ac To RS485/1 To CH 4/5 OpenSw OpenSw Linak sop-05 Fan CloseSw Soil temperature probe (Page 4.8) Soil water content CS616 (Page 4.7) TDR Pt100 RS485 WEST Systems Soil.
PAGE 2.14 On-Field Installation Accumulation chamber tubes Acc.Ch. RS485/1 RS485/2 RS485/3 Battery Aux.1 CH 4/5 CE CH 2/3 www.westsystems.com FL-I007 Red 6x4 Rilsan Tube Connect the accumulation chamber to the FluxStation respecting the InLet / OutLet. The InLet must be protected with the 2 inches filter mounted on the accumulation chambe base. Black 6x4 Rilsan Tube Filter Continuous monitoring device handbook www.westsystems.com Release 10.
3.1 Pump Brombach Outlet ASF THOMAS ASF THOMAS > Pump > Brombach Inlet Pump connections Pin 1 2 3 4 Color Brown White Blue Black Function +12V n.c. n.c. Gnd Pump specifications Rated flow: 1000 SCCM (Standard cubic centimeter per minute) Power supply 150 mA @ 12 Volts Manufacturer KNF, WEST Germany NMP830KNDC Pneumatic fittings diameter : 4 mm DC/DC ASF THOMAS ASF THOMAS 9..12/12V + WEST Systems Pump Continuous monitoring device handbook www.westsystems.com Release 10.
PAGE 3.2 Pneumatic circuit Pump maintenance ! ! ! Disassemble the pumping head Clean the diaphragm, the washer, and the valves using a compressed air flow. Reassemble the head. ASF THOMAS ASF THOMAS V Unscrew the 4 screws Pump V Brombach Screws (4) Pump head cover Pump head gasket Pump head body, with valves Pump diaphragm ASF THOMAS ASF THOMAS Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 The pump efficiency can be affected by deposits of dust or water.
3.3 P1 P2 Pump Brombach > > AWM3300V A B Filter Inlet ASF THOMAS ASF THOMAS Outlet Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 PAGE Pneumatic circuit ! To avoid problems on the reading of P.Flow, that can be affected by the new KNF pump efficiency, please adopt this schema in the pneumatic connections. WEST Systems All the internal connections are made with silicon tube 8 mm external diameter 3.
PAGE 3.4 Pneumatic circuit The fittings WEST Systems Tube fitting for RILSAN 6x4 and Silicon tubes M10 x 1mm WEST Systems Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 2.8 mm Diameter 25 mm Material : Chromed brass 44 mm The tubes 3 types of tube are used to realize the gas sampling line: !RILSAN tube 6 mm external diameter 4 mm internal diameter; ! SILICON tube 8 mm external diameter 3.3 mm internal diameter; ! SILICON 12.5 mm external diameter 4.
4.1 PAGE Sensors Carbon dioxide detector DRAGER Polytron IR CO2 Specifications !Accuracy 3% of reading for > 350 ppm !Signal noise: 1% of reading pk-pk noise Out 4~20 mA linearized. !IP63 Case with anti spray protection IP65. !Operating temperature : between -20 and +65 °C. !Air relative humidity 0~95%. Release 10.0 November 2007 !Power supply 14-24 Volt DC Typical 300mA @ 24V DB 9 Male : Ch.0 Channel CO2 Ch. Name CO2 ppm Ch.
Sensors Carbon dioxide detector internal connections K1 K2 K3 K4 Drager Cable Pin Color K1 Red K2 Blue K3 Yellow K4 Green Function DB9 pin +24V 9 Gnd 5 S.Low 7 S.High 2 Carbon dioxide detector setup For the setup of the correct full scale please refer to pages 56-57 of Polytron transmitter IR CO2 handbook and to page 3.4 of this manual.
4.3 PAGE Sensors Barometric pressure gauge PT100B Barometric pressure B.Pr. HPa 1 PRESSURE TRANSMITTER MODEL SERIAL NO RANGE OUTPUT SUPPLY PTB100A R47xxxxx 800-1060 hPa 0-5 VDC 10~30VDC EXT.TRIG AGND GND SUPPLY VOUT 9 8 7 6 5 4 3 2 1 DB 9 Male : Ch.1 Vaisala Cable Pin Color Supply Red Gnd Blue AGnd Yellow VOut Green Function DB9 pin +12V 4 Gnd 5 S.Low 7 S.High 2 WEST Systems Release 10.0 November 2007 Channel Ch. Name CH. # Continuous monitoring device handbook www.westsystems.com Release 10.
PAGE 4.4 Sensors Gas line flow meter Flow direction (P1 to P2) P1 P2 A B AWM3300V dP Channel Manufatcurer : WEST Systems (Sensor made by Honeywell-SensorTechnics) Ordering code :AWM3300 Gas flow Range 0 to 1300 SCCM Accuracy 2% of F.S. Power supply 10 Volt Typical 30 mA @ 24 Volt WESTSystems Systems WEST P.flow and dP Channels The P.Flow channel measures the gas flow in the sampling line. The value must be in the range 700-1450 SCCM. The dP channel measures the depression of the pump.
Manufatcurer : Campbell Scientific Ordering code : . . . CS616 / CS615 (Out of stock) Water content Range 5-50 % Accuracy 3% Release 10.0 November 2007 The CS616 (CS615) Water Content Reflectometer measures the volumetric water content of porous media using time-domain measurement methods. The Water Content Reflectometer consists of two stainless steel rods connected to a printed circuit board.
Sensors Soil temperature Manufatcurer : RS Ordering code :RS 321-6090 Type : Pt100 DIN IEC 751 Class B 4 wires Range 0 to 200 °C Max operating temperature 200°C Accuracy 0.3°C Probe length 300 mm Probe diameter 6 mm Cable length 3 Meter Cable PTFE and silicon insulated 3+T Pin Pin Pin Pin Male connector (to the SoilBox) 1 : RED 2 : RED 3 : White T : White WESTSystems Systems WEST 300 mm The meteorological sensors are described in the Appendix A. Release 10.
The flux is proportional to the concentration increase ratio ppm/sec. The proportionality factor depends on the chamber volume/surface ratio as well as the barometric pressure and the air temperature inside the accumulation chamber. 5.1 PAGE The accumulation chamber Time Concentration increase ppm/sec The station, in the default configuration record and store the barometric pressure, the air temperature and the soil temperature.
PAGE 5.2 The accumulation chamber The accumulation chamber factors 86400 P 10 6 R Tk V A Where ! ! ! ! ! P is the barometric pressure expressed in mBar (HPa) R is the gas constant 0.08314510 bar L K-1 mol-1 Tk is the air temperature expressed in Kelvin degree V is the chamber net volume in cubic meters A is the chamber inlet net area in square meters. The dimensions of the A.c.K.
The accumulation chamber -10 0.499 0.484 0.470 0.456 0.443 0.430 0.417 0.405 0.393 0.382 0.371 0.360 0.349 0.339 0.329 0.320 0.310 -5 0.489 0.475 0.461 0.448 0.435 0.422 0.410 0.398 0.386 0.375 0.364 0.353 0.343 0.333 0.323 0.314 0.305 0 0.480 0.466 0.453 0.440 0.427 0.414 0.402 0.390 0.379 0.368 0.357 0.347 0.337 0.327 0.317 0.308 0.299 5 0.472 0.458 0.445 0.432 0.419 0.407 0.395 0.383 0.372 0.361 0.351 0.340 0.331 0.321 0.312 0.302 0.294 Air.T. °C 10 15 0.463 0.455 0.450 0.442 0.437 0.429 0.424 0.
-10 21.94 21.30 20.68 20.07 19.49 18.92 18.37 17.83 17.31 16.80 16.31 15.84 15.37 14.92 14.49 14.06 13.65 -5 21.53 20.90 20.29 19.70 19.12 18.57 18.02 17.50 16.99 16.49 16.01 15.54 15.09 14.65 14.22 13.80 13.40 WEST Systems Bar. Pressure Hpa (mbar) 1013 983 955 927 900 873 848 823 799 776 753 731 710 689 669 649 630 WEST Systems 0 21.14 20.52 19.92 19.34 18.77 18.23 17.69 17.18 16.67 16.19 15.71 15.26 14.81 14.38 13.96 13.55 13.15 15 20.04 19.45 18.88 18.33 17.80 17.28 16.77 16.28 15.81 15.34 14.
C02 4.8 ppm, 85 s 2500 ppm Flux:50.6 ppm/s , r 0.999 The "perfect" curve 5.5 PAGE The accumulation chamber For clearness only the CO2 flux curve is shown. The shape of the curve is quite perfect and the computation of the flux is done with a very good accuracy: r (regression quality factor) very close to 1. 200 ppm 100 In the following figures some examples of strange shape flux curves: C02 4.8 ppm, 85 s 2500 ppm Flux:32.6 ppm/s , r 0.
WEST Systems 04 T- Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 sot-07 br-05 1_rf 2_rf WEST Systems EN VV VV NE -T 40 Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 480 sop-05 PAGE 5.
The accumulation chamber PAGE 5.7 Cable AC1 Color Black White Blue Yellow Red Brown Amph Pin 1 2 3 4 5 6 Cable AC2 Bulgin 9 pin male Cable AC2 Bulgin Pin 1 2 3 4 7 8 9 Amph 3+T pin female Color Red Blue n.c. n.c n.c Green Yellow Amph Pin 1 2 Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 Cable AC1 Bulgin Pin 1 2 3 4 5 6 Amph 6+T pin female T 3 WEST Systems WEST Systems Continuous monitoring device handbook www.westsystems.com Release 10.
SA1 M1 J9 7 6 5 4 3 2 1 F_LINAK J10 7 6 5 4 3 2 1 F_FAN J2 7 6 5 4 3 2 1 FAN FAN+ R3 1K D2 FAN BYT03 D1 R1 1K R2 1K EnB D3 D4 AC_Blue AC_Brown Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 EV6 FAN+ VDC EnBlue EnBrown FAN+ VDC EnBlue EnBrown EV6 EnBlue L2 10W 12V RS579219 Continuous monitoring device handbook www.westsystems.com Release 10.
Type: inductive proximity; Model: IM020BM60TB8 Range: 2mm- Shielded; Protection grade: IP67; Made by Wenglor Thread: M8 LED Light - On/Off IM020BM60TB8 real size Cables: Type: Nano change; Part #:403001I03M020 Brand: Brad Harrison PAGE WEST Systems WEST Systems Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 Two proximity switches control the full opening and full closing of the accumulation chamber. 5.9 Continuous monitoring device handbook www.westsystems.
The accumulation chamber PAGE 5.10 Acc.Ch. RS485/1 RS485/2 RS485/3 Battery Aux.2 CH 4/5 Aux.1 CE Cable AC2 CH 2/3 www.westsystems.com FL-I007 Cable AC1 To Aux 1 To Acc. Ch. CloseSw Fan Aux Linak OpenSw Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 Continuous monitoring device handbook www.westsystems.com Release 10.
Flux Station configuration/testing Palmtop (PDA) SetStation application Run SetStation on the PDA, and connect the PDA to the station following the instruction below. PAGE 6.1 Now turn on the FluxStation, pin-007 PTB100A R47xxxxx 800-1060 hPa 0-5 VDC 10~30VDC MODEL SERIAL NO RANGE OUTPUT SUPPLY PRESSURE TRANSMITTER EXT.TRIG AGND GND SUPPLY VOUT www.westsystems.
Flux Station configuration/testing Palmtop (PDA) SetStation application SetStation 1.2 Select : COM1 Serial Cable on COM1 Communication port Connected Station IP Status Station Date Get All Set All File Parameters SetStation 1.2 If the communication is successful the information on the screen will be updated. Communication port > COM1 Serial Cable on COM1 Status: Idle when the station is in StandBy / running when the station is performing a mesurement clicle; 0.
Flux Station configuration/testing Display readings form Now the infromation channels is displayed. SetStation 1.2 of the analogical PAGE 6.3 Digital CO2 ppm 350.45 1013.45 Ch.2 mA 0.00 Ch.3 mV 0.00 Ch.4 mV 0.00 Ch. 5 mV 0.00 dP. mBar 1013.45 P.Flow SCCM 4.05 Pump On RS485 scan AC Up Format SD Set Clock AC down SD Status Radio On Sample now SetStation 1.2 The readings are updated every second; Digital >> TDR % 50.45 TDR raw uSec 134.45 6B13 °C 32.00 0.
Flux Station configuration/testing Display readings form SD memory card commands. SetStation 1.2 Digital CO2 ppm 350.45 Bar.P.mBar Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 ! SD Ok 1013.45 Ch.2 mA 0.00 Ch.3 mV 0.00 Ch.4 mV 0.00 Ch. 5 mV 0.00 dP. mBar 1013.45 P.Flow SCCM By pressing [SD Status] an short report about the SD card appear; The “SD Card Ok” means that the station found the SD card and that’s working fine. 4.
Flux Station configuration/testing 6.5 Flux analisys parameters SetStation 1.2 Select the {Parameters}{Flux menu voice. Communication port idle refresh Station Date 06/12/07 12:34:36 Flux Analisys Data Logger readings GetDisplay All Set All Program timings Channels parameters File Parameters ok SetStation 1.
Flux Station configuration/testing Analisys every time It's the interval between analisys. Normally the station makes an analisys every hour and then the default value of the parameter is 3600 sec; Valid range 300 .. 7200 seconds; If the time between analisys is 3600 secs (1 hour) the analisys will start exactly at the beginning of hour, for instance the sampling time will be 8 sharp, 9 sharp ect.
Flux Station configuration/testing 6.7 PAGE Palmtop (PDA) SetStation application Set RADIO On/OFF: RADIO Off time (sec) The telemetry system will be turned OFF after every analisys starting from the time,expressed in minutes, you set. Green sector : the station is making the analisys Radio ON time: 600 sec 0 55 50 5 45 20 40 15 Continuous monitoring device handbook www.westsystems.com Release 10.
Flux Station configuration/testing Palmtop (PDA) SetStation application Delta T.: Is the time , expressed in millisecond, between the points sampled in the flux record. Normally the value is between 500 (0.5 seconds) and 3000 ( 3 seconds) The global flux record time is : Delta T. x Num. Samples/ 1000 CO2 ppm In the following plot a sample of flux record with Num of Samples = 32 and DeltaT. = 1000 (1 second). The overal flux record lenght is 32 second.
Flux Station configuration/testing 6.9 Data Logger parameters Communication port idle 0.1 refresh Station Date 06/12/07 12:34:36 Flux Analisys Data Logger readings GetDisplay All Set All Program timings Channels parameters File Parameters The fluxstation can works as Data Logger, but in this modality it don’t measure the flux, then this modality is not described here. Program Timings SetStation 1.2 Select the {Parameters}{Program Timings} menu voice.
PAGE 6.10 Flux Station configuration/testing Channels parameters Channels parameters Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 idle refresh Station Date 06/12/07 12:34:36 Flux Analisys Data Logger readings GetDisplay All Set All Program timings Channels parameters File Parameters You’ve to access this menu when you’ve to change: 1) the full scale value of the DRAGER CO2 detector; 2) the analog channels configuration; 3) the number of flux curves.
Channels parameters Name: is a description of the channel, to help the operator. 6.11 PAGE Flux Station configuration/testing FF06: the output of sensor connected to the channel is 4-20 mA and the fluxstation will consider 4mA equal to the “Low Scale” value and the 20 mA equal to the “Full Scale” value. FF05: the output of sensor connected to the channel is 0-5V and the fluxstation will consider 0 V equal to the “Low Scale” value and the 5 V equal to the “Full Scale” value. FF0F: special Conf.
Changing the Drager Polytron IR CO2 full scale value. Select the Channel #0 ok SetStation 1.2 Get completed! Name CO2 ppm Get Set Conf (Hex) FF06 Get Set Gain 1 Get Set Offset 0 Get Set Full scale 20000 Get Set Low scale 0 Get Set Get Set Track On Off Get All << Previous Ok Set All Next >> ok SetStation 1.
To change the full scale value of the carbon dioxide detector turn on the detector by running SetStation from the PDA. PASSWORD? 2 Press the enter key, then the display will ask to enter the password Press the DN key twice to set the password = 2 then press enter to confirm MENU: CONFIGURATION: Now press the DN key twice the display will show MENU: CONFIGURATION: MENU: CONFIGURATION: Now press the enter key to enter in the configuration menu CONFIGURATION: GAS CONFIGURATION CONFIG.: 2.
Changing the Drager Polytron IR CO2 full scale value. PAGE 6.14 Press Enter key to proceed ! Press Esc twice and Enter to come back at the measurement. Remember now to set the same value on the flux station, if not already done Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 CONFIGURATION: GAS CONFIGURATION WEST Systems WEST Systems Continuous monitoring device handbook www.westsystems.com Release 10.
Running FluxStation 7.1 PAGE Quasi-continuous monitoring Pump A B Power A AWM3300V CH3 WEST RS232 CPU CPU RUN MMC-SD CH2 WEST CH0 POWER ON PULL AND SWITCH POWER OFF ON +12 +24 To turn on the FluxStation Pull up and switch on the power switch.. After a quick boot the FluxStation starts the quasi-continuous monitoring task. 8:00 9:00 10:00 nd Open acc.chamber time Flux record See next pages Radio OFF 2 Line Cleaning time Pump OFF Radio ON Close acc.
PAGE 7.2 Running FluxStation Quasi-continuous monitoring pin-007 PTB100A R47xxxxx 800-1060 hPa 0-5 VDC 10~30VDC MODEL SERIAL NO RANGE OUTPUT SUPPLY PRESSURE TRANSMITTER Pump Power A P2 A CPU RUN Led light B AWM3300V CPU CH3 WEST RS232 CPU RUN MMC-SD CH0 WEST CH1 POWER POWER ON PULL AND SWITCH POWER OFF ON +12 +24 When the FluxStation runs the CPU RUN light flashes every second. . The FluxStation stores the result of analisys cicles on a SD memory card.
Running FluxStation 7.3 PAGE Quasi-continuous monitoring CPU RUN MMC-SD Dn: Write disabled CPU WEST Release 10.0 November 2007 Up: Write enabled CPU RUN MMC-SD WEST Systems WEST CPU Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 Before to insert the SD memory card in the CPU slot verify that the Write On/Off swicth of the card is in the enabled position.
PAGE 7.4 Running FluxStation Quasi-continuous monitoring After inserting the SD card check with SetStation (Page 6.4) that the FluxStation recognized the SD card. ! WESTSystems Systems WEST In order to inform the operator of the next writing operation, the CPU run led light will be solid ON for a 60 second period before start to write on the card.
Running FluxStation 7.5 PAGE Station status pin-007 Pump P2 A B Power A AWM3300V CH3 WEST RS232 CPU CPU RUN MMC-SD CH2 CPU RUN Led light CH1 CH0 WEST POWER POWER ON PULL AND SWITCH POWER OFF ON +12 +24 2 Line Cleaning time Radio OFF The station states: has two working 1) StandBy: the time between the analisys cicles; nd Flux record See next pages Open acc.chamber time Pump OFF Radio ON Close acc.
WESTSystems Systems WEST Release 10.0 November 2007 Continuous monitoring device handbook Release 10.0 November 2007www.westsystems.com PAGE 7.
General discussion The subjects regarding calibration of your instrument for the measurement of diffuse flux will be discussed in this chapter. 8.1 PAGE Calibrations As explained previously the flux measurement is proportional to the slope of the concentration curve versus time. The proportionality factor depends on the volume/surface ratio of the accumulation chamber used for the measurement, as well as, the barometric pressure and air temperature at the moment of making the measurement.
PAGE 8.2 Calibration of Drager IR CO2 When you need to calibrate the Drager The Drager is a quite stable detector, the thermal stabilization features allows the instrument to maintain it calibration for long periods. The calibration is necessary only when there is a drastical change in the barometric pressure. What you need to calibrate the Drager ! ! A cylinder of nitrogen or synthetic UPP Air, or a SODA LIME trap or a DRAGER PIPE to trap the carbon dioxide.
8.3 PAGE Calibration of Drager IR CO2 PASSWORD? 2 Press the DN key twice to set the password = 2 then press enter to confirm www.westsystems.com www.westsystems.com Press the enter key, then the display will ask to enter the password MENU: MEASUREMENT: Now press the enter key to enter in the maintenace menu MAINTENANCE: CALIBR.
PAGE 8.4 Calibration of Drager IR CO2 VALUE CONSTANT? 0.00 Y/N VALUE CONSTANT Step NEW CALIBRATION FACTOR STORED! Press the Enter key to proceed. VALUE OK? 0.00 www.westsystems.com www.westsystems.com Wait that the reading is stable then select Y by pressing the left arrow key then press the enter key. Y/N Y/N Select Yes to finish the ZERO calibration procedure MAINTENANCE: CALIBR. ZERO Press the DN key to select the calibration sensitivity menu voice MAINTENANCE: CALIBR.
Be sure that the TEDLAR bag filled with the CO2 standard gas mixture (18100 ppm of CO2) is connected to the inlet port of instrument and Start the pump of instrument by pressing the START button on PalmFLUX. Then press the enter key to enter in the SPAN calibration procedure. Remember that using a 5 liter TEDLAR bag you have 4-5 minutes to complete the procedure. Portable flux meter handbook Release 10.0 5.39 December 2007 2005 VALUE CONSTANT? 1.
Calibration of Drager IR CO2 PAGE 8.6 STOP GAS FLOW? Y/N VALUE < ALARMLIMIT? 1.80 Y/N Select Yes end press Enter to finish the span calibration procedure. www.westsystems.com www.westsystems.com Select Yes and press Enter to proceed. Portable flux meter handbook Release 10.0 5.39 December 2007 2005 Portable flux meter handbook Release 10.0 December 2007 WEST Systems WEST Systems Now the DRAGER detector is calibrated.
This calibration procedure can be necessary if the reading odf CO2 concentration on the FluxStation and the reading of the Drager are different. Normally there is a small difference between the DRAGER and FluxStation; This is due : !to the resolution of the display of Drager: +/- 10 ppm and +/- 100 ppm if the full scale is more than 10000 ppm; ! to the analog to digital conversion; ! to the leak of linearity of DRAGER in some ranges.
Calibration of the 4-20 mA interface of Drager PAGE 8.8 NEW CALIBRATION FACTOR STORED! Press Enter key to proceed Portable flux meter handbook Release 10.0 5.39 December 2007 2005 200 mV REAL VALUE? 200.0 Now the Drager shows: 200 mV REAL VALUE? In this phase the reading of concentration in the palmflux must equal to the full scale you set on the palmflux, more or less a tolerance value and a little noise. If the value you read is not correct you can adjust it.
Portable flux meter handbook Release 10.0 December 2007 Portable flux meter handbook Release 10.0 December 2007 WEST Systems WEST Systems www.westsystems.com www.westsystems.com PAGE 8.
Calibration of the flux meter Fluxmeter calibration As the computation of flux is a function of the gradient of concentration vs. time, barometric pressure, air temperature and the the accumulation chamber volume / surface ratio, calibration of the instrument depends only on the detector calibration.
8.11 PAGE Calibration of the flux meter Fluxmeter calibration For fluxes between 300 moles/(m2day) and 700 moles/ (m2day) the injected flux is controlled by means of a mechanical flow reducer and measured using a bubble flowmeter (Accuracy 3%) before and after the flux measurement with Two series of measure were performed for flux of 300 moles/(m2day) to evaluate the coherence between the two different methods of flow measurement.
PAGE 8.12 Calibration of the flux meter The influence of pumping flow Influence of pumping efficiency on the flux measurement results: We have carried out some sets of measures utilizing the same injected flux but with different pumping flow from the accumulation chamber to the detector. measured with a Microbridge Mass Airflow sensor (Accuracy 2%). We have not noted a significant variation of the measures except when the pumping flow is less than 200 SCCM. www.westsystems.com www.westsystems.
Calibration of the flux meter The influence of mixing device PAGE 8.13 In the plot the measured flux vs the mixing device rotation speed. Influence of the mixing device rotation speed on flux reading 0.5 0.4 0.3 0.2 0.1 0 -0.1 -0.2 -0.3 -0.4 -0.5 0 20 40 60 80 100 120 www.westsystems.com Influence of the mixing device rotation speed Error % www.westsystems.com The mixing device was suspected, by some researcher, to affect the accuracy of the flux measurement.
Portable flux meter handbook Release 10.0 December 2007 Portable flux meter handbook Release 10.0 December 2007 WEST Systems WEST Systems www.westsystems.com www.westsystems.com PAGE 8.
Databases structure The databases All the information necessary to run scada, as well as the data acquired by the fluxstation, are stored in a single file : scada_data.mdb PAGE 9.1 These two file are located in the C:\WEST\SCADA\ folder The two databases are accessed by our software via ODBC, then two System DSN (Data source name) definitions are added by the Scada setup. The first system DSN is scada_data , it has a MS Access (.mdb) driver and it point to the file C:\WEST\Scada\scada_data.mdb.
Databases structure PAGE 9.2 ! Both DSN are created automatically by the Scada suite setup. Normally it’s not necessary to modify it manually. Database scada_log.mdb The database scada_log.
Databases structure Scada_data.mdb is composed by several tables, some of them are to define the fluxstations and their configuration. Some of them contain the results acquired by the fluxstations.
Databases structure Database scada_data.mdb Each FluxStation must be defined in the Stations table.
WEST Systems Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 ! Type and Model are critical parameters that indicate to Scada how to manage the station and must be not modified. StationID is the station identifier and cannot to be miodified. The StationIP is the logical address of the station and cannot to be modified. PAGE The Site, Latitude, Longitude, Elevation and Responsible fields contain only descriptive data. WEST Systems Database scada_data.mdb 9.
Databases structure Database scada_data.mdb The PhisicalChannels table contain information about the “phisical” configuration of the station channels. ID 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 StationID OVK01 OVK01 OVK01 OVK01 OVK01 OVK01 OVK01 OVK01 OVK01 OVK01 OVK01 OVK01 OVK01 OVK01 OVK01 OVK01 OVK01 Number 0 1 2 3 4 5 6 7 1 2 3 4 169 170 171 172 173 Name FullScale LowScale Gain Offset Statistics PlotCurve Conf Type CO2 20000 0 1 0 1 -1 FF06 ANALOG Bar.P 1060 600 1 0 1 0 FF05 ANALOG Ch.
Database scada_data.mdb Channels table 9.7 PAGE Databases structure Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 Each record define a channel , for instance Air temperature or CO2 Flux. Most of the field of the table must not modified by the user and are not documented here.
PAGE 9.8 Databases structure Database scada_data.mdb Telemetry table DISK The station has not a telemetry system and the data is trasferred via SDmemory card. In the telemetry table we defined a telemetry of this base type: TelemetryID = SD WindowStart=00:00:00 WindowStop=23:59:59 CommPort= 0 HostID= Settings= TelemetryType=DISK BaseType=DISK CommTimeOut= If you want to set this telemetry for your station you’ve to put the Stations.
Database scada_data.mdb 9.9 PAGE Databases structure RS232 The base type RS232 supports several TelemetryTypes: Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 SSR24: Spread spectrum 2.4 or 0.9 GHz in switched mode; RS232: Direct cable connection. The base type RS232 supports several TelemetryTypes: The configuration of each type is described in the telemetry handbook.
WEST Systems WEST Systems Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 PAGE 9.
Scada software suite General description The software suite of Scada is composed by 4 applications: PAGE 10.1 Monitoring: This application allows you to have a “quick view” of the data acquired by the FluxStations. Query: This application allows you to query the database and obtain a table in the .xls format of the data stored in the database. Scada Scada reads the network configuration from the scada_data ODBC data souce name (Chapter 9).
PAGE 10.2 Scada software suite Scada application Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 PC Clock ! On the left-top corner of Scada the pc clock is shown: Scada will use this date/time information to set the clock of each station managed every time it will be able to talk with the station. Then to avoid confusion in the data you get you must check that the PC clock is correctly set.
Scada software suite 10.3 PAGE Scada application By selecting {Options}{Show advanced menu} an advanced station menu will be enabled, then now right clicking the StationID field a more complete menu is shown. {Sample Now} By selecting {Sample Now} you force the immediate execution of an analisys cicle. The command will be sent to the station during the next call, then to have an immediate effect {Sample Now} must be followed by {Call Now}.
PAGE 10.4 Scada software suite Scada application Right click Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 By right clicking on the Channel.Name field of a station a menu appears. Channel Info Configure Alarms Show graph {Channel Info} By selecting {Channel Info} a report of the channel configuration and status will be shown. WEST Systems {Configure Alarms} By selecting {Configure Alarms} will be possible to set alarm level and modalities for the channel.
Scada software suite Scada application The StationID field backcolor and forecolor change in function of the status of the call. PAGE 10.5 OKV01 OKV01 OKV01 OKV01 Black, Yellow: Scada was able to reach the station, but the station is not answering correctly; OKV01 OKV01 OKV01 OKV01 OKV01 OKV01 Blue, White: Scada was able to reach the station and get the data; This mus be the “normal” situation when Scada is waiting the next call.
WEST Systems Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 PAGE 10.
Scada software suite 10.7 PAGE Scada application The alarms For each channel is possible to set an alarm: DN thresholds All thresholds UP Alarm UP PreAlarm DN PreAlarm DN Alarm The alarm management will modify the backcolor of the result field of the channel in order to highlight when a value is not “correct” OKV01 350.0 ppm OKV01 350 ppm OKV01 850 ppm OKV01 OKV01 0.23 moles/sm/day 0.234 The channel result field backcolor can be: WHITE : if the channel has not alarms configurated.
WEST Systems Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 PAGE 10.8 Scada software suite Scada application In the lower right corner a dialog shows the “binary” communication that flows on the coomunication stream between computer and telemetry device: Some of the byte are “translated” in a visible format: for example the byte &0D (That correspond to the Enter key ) is diplayed as . The format is: dd/mm/yyyy hh:mm:ss : 12/12/2007 16.28.
Scada software suite Flux revision Fluxrevision allows the checking of flux curves acquired by the station. PAGE 11.1 Select your station. By selecting “Never revised” you can select the curves you never checked Select the gas specie (If the station has additional gas detectors) Number of selected curves / Total number of curves in the query period. The list of flux curves that correspond to your selection WEST Systems By selecting [Filter] you can apply a more restrictive filter to your query.
Scada software suite To see the flux curve click on the list. The flux curve On the right part of the form is reported the information related with the flux curve. The interval used to compute the regression is highlighted in red. The regression computed is drawn in yellow. The limit of the interval are the two red vertical lines. Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 Flux revision WEST Systems WEST Systems Continuous monitoring device handbook www.
Manual revision Modify the limits by dragging the vertical lines in the desired position. The flux, the slope and the ErrQ will be automatically updated. To save the new result right clik and select {Save Flux} Using the manual revision the computation of flux is completely under your responsability. You can select wrong interval and get wrong results, as in the drawing. The slope of the curve is positive, but the seleted interval, too short, gives a negative result.
Scada software suite Flux revision Automatic computation of flux. FluxRevision can compute the flux following 3 different algorithms: By selecting {Auto Best slope}: the flux will be computed searching for the higher flux; By selecting {Auto Best ErrQ}: the flux will be computed searching for the better regression quality; By selecting {Auto Best Product}: the flux will be computed searching for a compromise between higher flux and best regression quality.
Scada software suite 11.5 PAGE Flux revision Batch processing Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 same consideration for the Revision date, the ErrQ and the Flux. WEST Systems WEST Systems Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 By clicking on [Sampling Date] you can order the list by date in ascending or descending mode .
WEST Systems WEST Systems Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 PAGE 11.
Scada software suite Monitor Monitor is a simple application designed to have a quick view of the last days of data. Monitor display the data of the channels that are active and Channels.Monitored > 0 and the Channels.MonitoringGroup = 1; PAGE 11.7 have the field To have two Monitor, one showing a group of channels and the second showing a second group of channels you’ve to: 1) Open the Channels table of scada_data.mdb , put the value 2 on the Channels.
WEST Systems WEST Systems Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 PAGE 11.
Query The Query application allows you to exctract the data from the database and to put it an a table in the MS Excel format 11.9 PAGE Scada software suite Set the date interval you want to query and Select the Channels then press [Next >>]; The buttons [Last month], [Last week] [Last 24 hours] will help you by setting a “standard” time interval. Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 Press the [Next >>] button.
Scada software suite A resume of the query is reported: The column “Records” shows how many record are present in the time interval you selected for each channel; By pressing [Export All] you’ll get a table with a row for each record (and a column for each channel, plus one for the date/time); By setting the “Number of points per day” field, for instance putting 4, and pressing [Export] you’ll get a table with the 4 row for each day.
A.1 PAGE AirBox Release II , General description The air box is an hardware/software interface that convert the signal of Air Temperature/ Relative Humidity gauge and Sonic wind speed / direction gauge to a digital format compatible with the DF1A soilflux station. www.westsystems.com www.westsystems.com The output of the air box is a RS485 with the standard WESTS Systems pin out. Since the AirBox is designed to be the last device in the bus the RS485 line is terminated with a 120 Ohm embedded resistor.
A.2 PAGE AirBox Connectors pinout GILL Connections PCB Pin 5) 8) 7) 8) 6) +12V Gnd RD Gnd TD www.westsystems.com Air RH & T Connections Panel 6+T Pin 2: Pin 3: Pin 5: Pin 6: female +5V Data Sck Gnd connector Red Yellow Blue Black PCB Pin 9) +5V 11) Data 10) Sck 12) Gnd Rain Gauge Connections Panel 3+T male connector PCB Pin Pin 2: SW_A 13) Pin T: SW_B 14) www.westsystems.
A.3 PAGE AirBox Gill sonic wind speed and direction gauge Speed !Range 0..60 m/sec !Accuracy : 4% !Resolution : 0.01 m/sec D17 R4 TZB12V 10R TMBAT46 R24 3.6V NiCd TZB5V 100n 100n U1 1K R6 R7 74AC14M Y1 C32 C33 L1 L2 L3 L4 WEST 1K D4 1M 100n R21 100n CRYSTAL J10 1K R9 R26 CRES2 D14 Rg C8 100n U4 C3 Systems 1 2 3 4 5 6 7 Airbox_II.
A.4 PAGE AirBox Silicon Air Temperature and relative humidity gauge SEN1 VDD SCK DATA GND SHT1X 4 3 2 1 VDD SCK DATA GND TMMBAT46 D1 R1 1K +5V SCKD DATAD 1K R2 1 2 3 4 5 6 7 D2 TZB 5V Probe Layout (actual size) 5 Sck 6 GND TZB 5VR1 1K 1 Sck Vdd Gnd Data AirT&RH SHT75 1K D2 R2 WEST 3 Data 2 +5V D1 Amphenol 6T male AirBox Two handbook Release 10.0 November 2007 J1 AMP 6+T MALE WEST Systems Probe schematics 4 November 2007 Release 10.0 WEST Systems AirBox Two handbook www.
AirBox Silicon Air Temperature and relative humidity gauge PAGE A.5 Air Temperature accuracy Vs. temperature 1.6 1.4 1.2 1 0.8 0.6 0.4 0.2 0 -40 -20 0 20 40 60 80 100 120 Measured temperature °C www.westsystems.com Error °C www.westsystems.com Temperature accuracy 4 3 2 1 0 0 20 40 60 %RH 80 100 120 AirBox Two handbook Release 10.0 November 2007 5 WEST Systems Relative Humidity Accuracy Error % November 2007 Release 10.
A.6 TZB5V 100n 7805 VR1 TH 510 www.westsystems.com www.westsystems.com D1 Multimeter probes 100n 100n ST232CD 2 +5V WEST R2 5 Sck 6 GND R24 D20 C4 100n C6 100n 100n C23 GILL C24 C25 SHT1X U1 8 R6 SEN2 Gnd 1 U2 1M 12) Ss Gnd JTAG C22 R25 Data S2 SW DIP-8 R27 J1 ATMEGA128/64 C2 11) Ss Data 100 C30 R7 J10 Y1 C33 L1 L2 L3 L4 C32 22p 22p WEST D4 D14 R21 100n CRYSTAL Pluvio 1K Rg CRES2 74AC14M 14) Rg Sw.B R9 R26 1M Rg 100n 13) Rg Sw.
Rain gauge PAGE A.7 Manufatcurer : LASTEM Ordering code :DQA030 Rain gauge Range: illimited . . . . Accuracy 1% Resolution 0.2 mm Collector surface area: 324 square centimeter Collector area diameter : 203 mm Temperature range 0..+50°C AirBox While the FluxStation is in stand-by state the airbox reads and store the reading of the raingauge. The airbox is power supplied with a 3.6V rechargeable battery that’s charged by the FluxStation every analisys cicle. Check www.westsystems.com www.westsystems.
Rain gauge PAGE A.8 levelling bolts In order to have a correct working the raing-gauge must be levelled by trimming the 3 levelling bolts. A spirit-level will help in the levelling. www.westsystems.com AirBox Two handbook Release 10.0 November 2007 spirit level WEST Systems Release 10.0 November 2007 Unscrew the 3 bolts WEST Systems AirBox Two handbook www.westsystems.
Cables PAGE C.
PAGE C.2 Cables Standard analog sensor cable WEST Systems Continuous monitoring device handbook www.westsystems.com Release 10.0 November 2007 3 meter Amphenol 6+T Female (sensor side) Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Pin 7 Amphenol 6+T Male (station side) 12 Volts Channel A signal Signal GND common Channel B signal n.c. Gnd (power ground) n.c.
Global system block scheme. The system described in this handbook is a data trasmission system able to transfer bidirectional data between two RS232 based equipments. F.1 PAGE SpreadSpectrum 2.4 GHz radio telemetry The system is composed by 2 radio-modems, 2 antennas, and several accessories. RS232C Master radio-modem 2.4 GHz SS Radio links RS232C FluxStation Slave radio-modem The maximum range of "radio link" is approximately 35 Km (20 miles) in optimal conditions.
PAGE F.2 SpreadSpectrum 2.4 GHz radio telemetry Point to point telemetry system The system is set to trasmit data bidirectionally from the master center up to the remote equipment: Due to the specification of the FluxStation the maximum effective connection rate is limited to 38400bit/sec. PC -- Master SSR24 radio: 38400 bit/sec; 8 data bit; 1 stop bit; None parity; None handshake.
SpreadSpectrum 2.4 GHz radio telemetry PAGE Point to point telemetry system RS232 cabling specifications Female 9 pin DB9 connector Male 9 pin DB9 connector WEST Systems Slave radio-modem WEST Systems Male 9 pin DB9 connector Radio link Freewave based point to point telemetry www.westsystems.com Release 10.0 November 2007 Female 9 pin DB9 connector SSR24 based point to point telemetry Release 10.0 November 2007 Master radio-modem Master Center Computer FluxStation F.
PAGE F.4 SpreadSpectrum 2.4 GHz radio telemetry Connecting telemetry to the station P1 P2 A B CPU RUN MMC-SD POWER POWER ON PULL AND SWITCH POWER OFF ON +12 +24 SSR24 based point to point telemetry Release 10.0 November 2007 WEST CPU WEST Freewave based point to point telemetry www.westsystems.com Release 10.
P1 P2 A F.5 PAGE Connecting telemetry to the station Fogo Station SpreadSpectrum 2.4 GHz radio telemetry B AWM3300V POWER ON PULL AND SWITCH POWER OFF +12 +24 Connect the antenna out of radio to the pass-trough connector RF Radio pig-tail 0.25 meter station-side antenna cable SMA Male We supplied two ruggedized BackFire 18 db antennas. The antenna can be polarized in both , vertical or horizontal polarization. By default the antennas are set to the vertical polarization.
PAGE F.6 SpreadSpectrum 2.4 GHz radio telemetry Connecting the antenna to the station Connect the coaxial cable to the passtrought connection on the right panel of FluxStation. The antenna is furnished with a tilt adapter to point the antenna in the vertical plane. Due to space problems the tilt adapter must be assembled by you. SSR24 based point to point telemetry Release 10.1 November 2007 Pay attention to antenna polarization (Vertical in the picture).
Setting the radio-modem Change the setting of the radio-modem only when strictly necessary. F.7 PAGE SpreadSpectrum 2.
SpreadSpectrum 2.4 GHz radio telemetry Setting the telemetry. First communication and fluxstation-computer synchronization; The FluxStation clock can be not synchronized with the computer one; then, since the radio on/off is managed by the station a synchronization is necessary. That can be done using the PDA and the SetStation application or via computer.
Setting the telemetry. Point to point switched with repeater F.9 PAGE SpreadSpectrum 2.4 GHz radio telemetry Enter in the setup of master radio and : Set the master radio mode to 6: master switched; Open the call book entry of the radio; Set the call book entry 0 by setting the 2427581 radio as slave to be connected, setting the correct sequence of repeater that must be used. Open the stations table on scada_data.mdb.
PAGE F.10 SpreadSpectrum 2.4 GHz radio telemetry The radio-modem LED Lights You can check the status of the link watching the LED lights on the radio-modem front panel. The meaning of the LED light depends on the function assigned to the radio-modem. The LED light status for the Master center radio is reported on the drawing below. CD TX When you turn on the master radio-modem the 3 lights are solid red until the radio-modem connect to the repeater and the slave radiomodem.
SpreadSpectrum 2.4 GHz radio telemetry PAGE The radio-modem LED Lights F.11 Slave LED Status CD When you turn on the Slave radio-modem the 3 lights are as shown on the right until it will be connected to the master radio-modem via repeater.
SMA Antenna connector 50 Ohm Power Supply inlet 10 .. 24 VDC, max 1 Ampere @12 VDC 1 2 4 3 WEST Systems SSR24 based point to point telemetry Release 10.0 November 2007 RS232 port DTE female www.westsystems.com RS232 459 PAGE SET CD TX CTS www.westsystems.com Freewave based point to point telemetry www.westsystems.com Release 10.0 November 2007 Status LED Lights ANTENNA 459 WEST Systems F.12 SpreadSpectrum 2.
SpreadSpectrum 2.4 GHz radio telemetry 1 2 3 4 PAGE The radio modem interface schematic F.13 Power IN 1N4007 J11 D9 1A +12V C16 100n D10 24V600W J7 1 2 3 4 5 6 7 8 9 10 R15 +VFW SET DTR_FW GND TDFW GND RDFW CD RTS CTS J10 RS232 FEMALE 1K 1 DSR 6 2 7 3 8 4 9 GND 5 S2 Set Btn SSR24 based point to point telemetry Release 10.0 November 2007 Radio JP2 1 2 D T R _F W DTR da PC The radio modem PCB layout J2 PS In +VDC GND Freewave based point to point telemetry www.westsystems.
F.12 SpreadSpectrum 2.
SoilBox General description The Soil.box is an hardware/software interface that convert the signal of Soil Temperature and of the Soil Water Content TDR CS616 sensor to a digital format compatible with the DF1A soilflux station. PAGE S.1 TDR 1 2 3 6 IN-A T Gnd-Black Gnd RS485A RS485B +12 V RS485 www.westsystems.com 1 Ign-Orange HUB-04 IN-B OUT Pt100 www.westsystems.com Pt100 TDR Ohm TDR P uSec s/n CE Soil WC % T. °C The output of the Soil.
PAGE S.2 SoilBox Connectors pinout Soil Temperature Pt100 Connections www.westsystems.com Panel 3+T Pin 1: Pin 2: Pin 3: Pin T: ! female connector Red 1 Red 2 White 1 White 2 If you’re using a old Pt100 probe , supplied with the Rs485 HUB check that the pin T is connected to the White 2 wire of the Pt100 probe. If NOT please modify the Pt100 connection. A simple bridge between pin 3 and pin T can be enough , also if this simple solution can reduce the temperature measurement accuracy.
1A F1 1K R18 B A Vp 1 2 3 4 5 6 7 CS61x J10 V12 U7 VN820 1N4007 D14 10R R33 10R R32 Vp C10 D15 100n TZB12V D17 D16 VCC T1OUT T1IN T2OUT T2IN R1IN R1OUT R2IN R2OUT ST232CD 6 7 7805 VIN D R DE RE 100n C31 VR2 OUT 12 9 11 10 1 3 4 5 VCC C11 100n VCC 8 4 1 3 2 D1 C22 100n R29 1M SCK MOSI MISO CS616EN VCC14 1 19 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 1 PE0 RXD0/(PDI) PE1 TXD0/(PDO) PE2 (XCK0/AIN0) PE3 (OC3A/AIN1) PE4 (OC3B/INT4) PE5 (OC3C/INT5) PE6 (T3/INT6) PE7
SoilBox PCB Layout 100n C33 C32 L6 C30 C25 6K19 0.1% 100n R39 C26 HUB2004.DDB + p 100n Red2 Red1 J9 Pt100 Input +12VS C43 CS61x 1K R21 C31 U10 BC817 White 2 Pin T White 1 Pin 3 Red 2 Pin 2 Red 1 Pin 1 Ignition Pin 1(Yellow) P.
SoilBox Soil water content PAGE S.5 6+T Pin Pin Pin Pin Pin Pin Pin Male cable connector (To the SoilBox) 1 : Red +12 Volt 2 : Orange Ignition 3 : Green Frequency signal out 4 : Black+Sh. Ground 5 : n.c. 6 : n.c. T : n.c. WEST Systems SoilBox handbook Release 6.9 October 2004 The CS616 (CS615) Water Content Reflectometer measures the volumetric water content of porous media using time-domain measurement methods.
www.westsystems.com PAGE S.6 SoilBox Soil temperature Manufatcurer : RS Ordering code :RS 321-6090 Type : Pt100 DIN IEC 751 Class B 4 wires Range 0 to 200 °C Max operating temperature 200°C Accuracy 0.3°C Probe length 300 mm Probe diameter 6 mm Cable length 3 Meter Cable PTFE and silicon insulated WEST Systems SoilBox handbook Release 6.
PDA software installation The SetStation software is furnished by WEST Systems on two different support format: !On the WEST Systems software CD ! On the Compact Flash data card. PAGE Z.1 4) Copy the file SetStation.exe file in the folder \\Mobile device\PocketPC\Windows\Start Menu "storage card" based installation ACER n300 WEST Systems acer n300 All Palmtops: !Insert the Compact Flash / Secure Digital Card in the palmtop; !Run "File Explorer" on your palmtop; !Select the file "SetStation.
Software installation File explorer My device www.westsystems.com iPAQ File store My Documents Program Files Storage card Temp Windows File explorer Storage card SetStation Cut Copy Rename Delete To select and copy the file in the "storage card": Tap "Start" , "Programs" then tap "File Explorer". From the "File explorer" menu tap "My Device" and then tap "Storage card" Now tap and hold SetStation until a flying menu appear. Tap "Copy" www.westsystems.com PAGE Z.
Scada software installation The Scada software is furnished by WEST Systems on software CD. the WEST Systems PAGE Z.3 Requirements OS Requirements The Scada software suite can run on MS Windows XP professional, MS Vista pro / business. We don’t advise other OS such as XP Home or Vista Home. First installation Run ScadaSetup.msi www.westsystems.com www.westsystems.com The Scada software suite requires the installation of the MS .NET Framework 2.0.
WEST Systems WEST Systems FluxStation handbook Release 10.0 December 2007 FluxStation handbook Release 10.0 December 2007 www.westsystems.com www.westsystems.com PAGE Z.
HandBook 10.0 released on November 2007 FluxStation HWR8 and Scada software suite HandBook 10.1 released on December 2007 Some minor revision were introduced: Chapter 2 : Assembly instruction Appendix F: Page F.6 antenna assembly instruction © www.westsystems.com Continuous monitoring Release 10.
W Sys ES tem T s © www.westsystems.com Continuous monitoring Release 10.