User manual
If the unit displays ‘Err2’ after Step
5, it means the excessive saliva went
into the sensor during the test. Wait
20 minutes and then try again.
When the total number of tests is over
500 times, the recalibration service
is needed. Please send the device
to ACE.
If the unit displays ‘CAL’ after
countdown, it means the tester has
been used over 365 days from its last
calibration. Please send the device for
a calibration service to ACE.
Remaining days to recalibration,
e.g. d365 means 365 days left to
recalibration. Contact ACE for proper
recalibration service.
The elapsed day from the required
recalibration date, which is a reminder
to do the recalibration, e.g. d.001
means 1 day elapsed from the requi-
red recalibration date. Please send
the device for a calibration service
to ACE.
Display Meaning/Action
Technical data
0.07 ~ 5.00 ‰ (fixed basic setting),
under 0.07 ‰ the test result is shown
as 0.00 ‰
+/- 0.05 ‰ at 0.5 ‰
15 seconds up to max. 2 minutes
Within 12 seconds
Fuel cell sensor
2 pieces AAA alkaline batteries (1.5 V)
Approx. 1.000 tests
47 mm x 108 mm x 17 mm (W x H x D)
85 g (including batteries)
Every 6-12 months or after approx.
500 tests
Operation: 5 - 40° C
Storage: 0 - 50° C
Statement of the BAC
Accuracy
Warm-up time
Response time
Sensor
Energy supply
Battery capacity
Dimensions of the device
Weight
Calibration
Ambient conditions