Data Sheet

Technical data*
Sensor
Premium fuel cell sensor
Accuracy
+/- 0.05‰ (per mill) with reference to 1.00‰ at 25°C (‰ = default setting)
OR +/- 0.005% BAC (Blood Alcohol Concentration) with reference to 0.100% BAC at 25°C
OR +/- 0.025mg/L with reference to 0.50mg/L at 25°C
Detection
Range
0.00 up to 5.00‰ (‰ = default setting)
OR 0.000 up to 0.500% BAC (USA)
OR 0.00 up to 2.50mg/L (UK)
Warm-up Time
Within 10 seconds at 0.5‰. This may vary depending on the value of per mill / BAC which was
measured
Test memory
Stores the results of up to 100 tests
(with test number, date and time)
Dimension
128 x 60 x 24mm (L x W x H)
Weight
130g (batteries included)
Power Supply
2 pcs. of 1.5V AA Alkaline batteries (power for up to 1,000 tests)
Working
Temperature
-5°C to 40°C
Storage
Temperature
-5°C to 40°C
Connectivity
(only with
optional BT)
Frequency range: 2402 to 2480 MHz
Maximum transmission power: 1,6 dBm
Calibration
Default setting: After 12 months or 500 tests (whichever comes first). The settings can be changed
(see chapter “Advanced Menu Settings”). Please observe the calibration instructions on the display.
You can send the instrument for calibration to the service address given at www.alkomat.net.
*The technical data are subject to change without prior notice for functional improvements.