2.8 MAINTENANCE AND CARE
Risk posed by voltage when performing work on the BluFlame
Muscle cramps, electrical burns, unconsciousness, respiratory
arrest, and death.
Maintenance and repair work on electrical equipment
should be performed exclusively by trained, qualified
personnel.
Check to make sure that the BluFlame is de-energized.
Maintenance work performed incorrectly
Bodily injury and property damage resulting from
unauthorized maintenance work
Maintenance work should be performed exclusively by
personnel briefed on all safety aspects.
Follow all safety instructions.
Risk of bodily injury and property damage posed by hot surfaces
When hot, the probe tube can cause burn injuries, as well as
fire damage when placed on a combustible surface.
Maintenance work should be performed exclusively by
personnel briefed on all safety aspects.
Follow all safety instructions.
Let the probe cool down after the measurement is complete.
Risk of bodily injury and property damage posed by caustic
liquid
The liquid emptied from the condensate trap may be
slightly acidic.
Follow all safety instructions.
Wear work gloves.
In case of skin contact, clean the affected areas IMMEDIATELY.
Make sure that your eyes do not come into contact
with the liquid.
Thoroughly clean all parts that come into contact with
the condensate.
2.9 ENVIRONMENTAL PROTECTION
During all work with the BluFlame, make sure to comply with
all applicable legal regulations concerning waste prevention
and the proper recycling and disposal of environmentally
hazardous substances.
Make sure that environmentally hazardous substances
(cleaning liquids containing solvents) do not contaminate
the ground or reach sewer systems.
Collect, store, and transport environmentally hazardous
substances in suitable containers and then dispose of them.
Device Description
The BluFlame measuring device can be used to obtain precise check and
adjustment measurements on gas-fired, oil-fired, and wood-burning appliances.
3.1 SCHEMATIC GAS SYSTEM DIAGRAM
In combination with a flue gas probe, the BluFlame sucks in part of the flue
gas from the combustion air duct and analyzes its components with the
use of electrochemical sensors. The pressure (draft) and temperature are
measured directly at the probe tip.
1. Flue Probe 4. Filter 7. O
2
Sensor
2. Condensate Trap 5. Check Valve 8. CO Sensor
3. Draft Measurement 6. Gas Pump 9. NO Sensor SKUA10773NO only