11.6 O-RING KIT FOR CONDENSATE TRAP
1. Outer gasket #65627
2. Inner gasket #65628
11.5 TROUBLESHOOTING
SIGN FAULT CAUSE SOLUTION
The BluFlame is not responding at all
The BluFlame is not responding to
touch controls
Press the Reset button
Excessive cooling inside the device;
the BluFlame cannot be used
Display prompt:
“Device too cold” or
beep every 5 seconds
The BluFlame was stored in an
excessively cold environment
Place the BluFlame in a warm room
and wait until it is ready
Incorrect readings Zeroing error
The sensors were already exposed to
gas during calibration
Purge the BluFlame with fresh air
and restart it
The BluFlame will not turn on or is
not responding after being turned on
No more battery charge
Connect the BluFlame to an outlet so
that the battery is charged
Measurement without accurate
temperature values
Temperature display:
- - - , - °C / °F
Faulty thermocouple, compensating
cable discontinuity or compensating
cable not connected
Call After-Sales Service.
Remove the probe from the flue pipe
and condensate from the probe tube
Incorrect gas readings
Measuring range exceeded:
O2 value too high, CO and CO2 values
too low
“Probe-device” connection not
established properly
Probe / tube / condensate trap leak
Pump not sucking correctly
Run a leak test – refer to section 9.4,
“Leak test,” on page 17
Leaky spots can be detected by
visually inspecting the probe, tubes,
condensate trap
Incorrect temperature readings Gas temperature too high or jumping
Probe connector not connected
correctly
Cable discontinuity in probe line
Condensate formation on probe tip
Check the probe connector and the
probe line for breaks (loose contact)
Shake off condensate from the
probe tip