User Guide

19
READINGS
ELECTROCHEMICAL SENSOR O
2
Measuring Range 0 ... 21%
Resolution 0.1 Vol. %
Abs. Accuracy ± 0.2 Vol. %
T90 Response Time < 30 s
Expected Service Life Years on Air 2
ELECTROCHEMICAL SENSOR
(H2 – COMPENSATED)
CO/H2
Nominal Measuring Range 0 ... 4000 ppm
Overload Range < 10000 ppm
Resolution 1 ppm
Absolute Accuracy /
% of Reading
± 10 ppm /
5% (0 ... 4000 ppm)
10% (> 4000 ppm)
T90 Response Time < 40 s
ELECTROCHEMICAL SENSOR
(NOT H2 – COMPENSATED)
CO
Nominal Measuring Range 0 ... 5000 ppm
Overload Range < 10000 ppm
Resolution 1 ppm
Absolute Accuracy /
% of Reading
± 10 ppm /
5% (0 ... 4000 ppm)
10% (> 4000 ppm)
T90 Response Time < 30 s
ELECTROCHEMICAL SENSOR
(NOT H2 – COMPENSATED)
CO (#65929)
Nominal Measuring Range 0 ... 10000 ppm
Overload Range < 20000 ppm
Resolution 1 ppm
Absolute Accuracy /
% of Reading
± 10 ppm /
5% (0 .. 10000 ppm)
10% (> 20000 ppm)
T90 Response Time < 30 s
ELECTROCHEMICAL SENSOR NO
Nominal Measuring Range 0 ... 5000 ppm
Overload Range < 10000 ppm
Resolution 1 ppm
Absolute Accuracy /
% of Reading
± 5 ppm /
5% (0 ... 1000 ppm)
10% (> 1000 ppm)
T90 Response Time < 50 s
TEMPERATURE MEASUREMENT T1, T2
Number of Type K Thermocouple
Inputs
2
Measuring Range -40 °C ... 1200 °C / -40 °F ... 2192 °F
Accuracy
±2 °C /±3.6 °F
0.50%
FLUE GAS TEMPERATURE
(WITH PROBE)
T
A
Measuring Range with Stainless
Steel Gas Sampling Tube
0 °C... 650 °C / 32 °F ... 1202 °F
Abs. Accuracy / of Reading
±2 °C /±3.6 °F
0.50%
COMBUSTION AIR TEMPERATURE
(WITH USER SUPPLIED K-TYPE SEN.)
T
I
Measuring Range with Combustion
Air Sensor
0 °C ... 100 °C / 32 °F ... 212 °F
Abs. Accuracy 1 °C / 1.8 °F
FLUE DRAFT
Measuring Range ± 200 hPa / 2.9 psi
Abs. Accuracy / of Reading
0.00029 psi
1%
DIFFERENTIAL PRESSURE
Measuring Range ± 200 hPa / 2.9 psi
Abs. Accuracy / of Reading
0.00029 psi
1%